Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number 115-066755-00
Enforcement
- Recall Number
- Z-2949-2024
- Event ID
- 95099
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Mindray DS USA, Inc. dba Mindray North America
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 11, 2024
- Initiation Date
- July 22, 2024
- Classification Date
- September 4, 2024
- Address
- 800 Macarthur Blvd, Mahwah, NJ, 07430-2001, United States
Description
Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number 115-066755-00
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
UDI-DI: 06936415975784 Serial Number: AH6D28000517, AH6D28000518
US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada
2 units