Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged and non-centrifuged sample tubes. The cobas p 612 is intended for use with analyzers that perform tests in the area of clinical chemistry, immuno chemistry, coagulation, hematology, urinalysis, nucleic acid testing.
Enforcement
- Recall Number
- Z-2944-2018
- Event ID
- 80617
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diagnostics Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 5, 2018
- Initiation Date
- July 16, 2018
- Classification Date
- August 27, 2018
- Termination Date
- May 16, 2019
- Address
- 9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States
Description
Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged and non-centrifuged sample tubes. The cobas p 612 is intended for use with analyzers that perform tests in the area of clinical chemistry, immuno chemistry, coagulation, hematology, urinalysis, nucleic acid testing.
Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible presence of biological material on the nozzle may cause contamination resulting in potential false positive results, depending on the sensitivity of the analytical technology.
All units
US Distribution to states of: CA, NY,TX, WA, and WI.
12 units