Talladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, Model No. 5147-A1;(UDI): D7915147A10 - Product Usage: are intended for CAD/CAM fabrication of all-ceramic (no metal) dental restorations. Luminesse ZR Blanks are partially sintered Yttria stabilized zirconia blanks for use as CAD/CAM milling blanks.
Enforcement
- Recall Number
- Z-2943-2020
- Event ID
- 86227
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Talladium Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- September 16, 2020
- Initiation Date
- August 6, 2020
- Classification Date
- September 9, 2020
- Termination Date
- October 7, 2022
- Address
- 27360 Muirfield Ln, Valencia, CA, 91355-1010, United States
Description
Talladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, Model No. 5147-A1;(UDI): D7915147A10 - Product Usage: are intended for CAD/CAM fabrication of all-ceramic (no metal) dental restorations. Luminesse ZR Blanks are partially sintered Yttria stabilized zirconia blanks for use as CAD/CAM milling blanks.
The linear shrinkage on the zirconia blank disc s part label is incorrect. The average linear shrinkage found on the product label was outside the acceptable range. The acceptable average linear shrinkage ranges between 1.225 to 1.275. The production lot being recalled has a linear shrinkage rate of 2.251 which is above the acceptable range.
Lot Number.: 1AZG
U.S. Nationwide distribution including in the states of CA, FL, KS, LA, MO, and PA. No OUS Consignees.
39 pieces