FDA Enforcement
Class II
Terminated
DUO FLOW CUSTOM TRAYS; TRAY #414 TRAY #415 TRAY #67
Recall: Z-2941-2018
·
Reported September 5, 2018
Enforcement
- Recall Number
- Z-2941-2018
- Event ID
- 80581
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medical Components, Inc dba MedComp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 5, 2018
- Initiation Date
- July 12, 2018
- Classification Date
- August 25, 2018
- Termination Date
- July 2, 2020
- Address
- 1499 Delp Dr, N/A, Harleysville, PA, 19438-2936, United States
Description
DUO FLOW CUSTOM TRAYS; TRAY #414 TRAY #415 TRAY #67
Reason
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Code Info
MCBC900 MCBP350 MCDC900 MCGA140 MCCD760 MKAL320 MCAS020 MCCF310 MCCV750 MCCV870 MCDL720 MCDV440 MCFJ760 MKAF680
Distribution
Nationwide distribution; worldwide distribution.
Quantity
N/A