RAPIDLab 1265 Blood Gas Analyzer; 110321852, 10470366, 10491395 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.
Enforcement
- Recall Number
- Z-2930-2018
- Event ID
- 80516
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 5, 2018
- Initiation Date
- July 16, 2018
- Classification Date
- August 24, 2018
- Termination Date
- June 23, 2020
- Address
- 500 Gbc Dr Ms 514, Po Box 6101, Newark, DE, 19702-2466, United States
Description
RAPIDLab 1265 Blood Gas Analyzer; 110321852, 10470366, 10491395 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.
Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for carboxyhemoglobin (fCOHb) and methemoglobin (fMetHb). A negative interference with fCOHb has the potential to alter the medical assessment of the patient and may withhold necessary follow-up treatment and/or initiate cessation of medication in response to elevated fMetHb levels.
All units affected
Nationwide distribution. Worldwide foreign distribution.
2357