StimQ Peripheral Nerve Stimulator System (PNS) Model FR4A-TRL-AO US - Product Usage: is used for peripheral neural stimulation to provide therapeutic relief for chronic, intractable pain of peripheral nerve origin. The therapy utilizes pulsed electrical current to create an electrical energy field that acts on peripheral nerves in the limbs and torso to alter the transmission of pain signals to the brain.
Enforcement
- Recall Number
- Z-2929-2020
- Event ID
- 86115
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Stimwave Technologies Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- September 9, 2020
- Initiation Date
- July 15, 2020
- Classification Date
- September 2, 2020
- Address
- 1310 Park Central Blvd S, N/A, Pompano Beach, FL, 33064-2217, United States
Description
StimQ Peripheral Nerve Stimulator System (PNS) Model FR4A-TRL-AO US - Product Usage: is used for peripheral neural stimulation to provide therapeutic relief for chronic, intractable pain of peripheral nerve origin. The therapy utilizes pulsed electrical current to create an electrical energy field that acts on peripheral nerves in the limbs and torso to alter the transmission of pain signals to the brain.
The product contains a non-functional component not referenced in product labeling.
Serial Numbers: 211662-1, 211662-10, 211662-11, 211662-12, 211662-13, 211662-14, 211662-15, 211662-16, 211662-17, 211662-18, 211662-19, 211662-2, 211662-20, 211662-21, 211662-22, 211662-23, 211662-24, 211662-25, 211662-26, 211662-27, 211662-28, 211662-29, 211662-3, 211662-30, 211662-31, 211662-32, 211662-33, 211662-34, 211662-35, 211662-36, 211662-37, 211662-38, 211662-39, 211662-4, 211662-40, 211662-41, 211662-42, 211662-43, 211662-44, 211662-45, 211662-46, 211662-47, 211662-48, 211662-49, 211662-5, 211662-50, 211662-51, 211662-52, 211662-53, 211662-54, 211662-55, 211662-56, 211662-57, 211662-58, 211662-59, 211662-6, 211662-60, 211662-61, 211662-62, 211662-63, 211662-64, 211662-65, 211662-66, 211662-67, 211662-68, 211662-69, 211662-7, 211662-70, 211662-71, 211662-72, 211662-73, 211662-74, 211662-75, 211662-76, 211662-77, 211662-78, 211662-79, 211662-8, 211662-80, 211662-81, 211662-82, 211662-83, 211662-84, 211662-85, 211662-86, 211662-87, 211662-88, 211662-89, 211662-9, 211662-90, 211662-91, , 211613-1, 211613-10, 211613-100, 211613-101, 211613-102, 211613-103, 211613-104, 211613-105, 211613-106, 211613-107, 211613-108, 211613-109, 211613-11, 211613-110, 211613-12, 211613-13, 211613-14, 211613-15, 211613-16, 211613-17, 211613-18, 211613-19, 211613-2, 211613-20, 211613-21, 211613-22, 211613-23, 211613-24, 211613-25, 211613-26, 211613-27, 211613-28, 211613-29, 211613-3, 211613-30, 211613-31, 211613-32, 211613-33, 211613-34, 211613-35, 211613-36, 211613-37, 211613-38, 211613-39, 211613-4, 211613-40, 211613-41, 211613-42, 211613-43, 211613-44, 211613-45, 211613-46, 211613-47, 211613-48, 211613-49, 211613-5, 211613-50, 211613-51, 211613-52, 211613-53, 211613-54, 211613-55, 211613-56, 211613-57, 211613-58, 211613-59, 211613-6, 211613-60, 211613-61, 211613-62, 211613-63, 211613-64, 211613-65, 211613-66, 211613-67, 211613-68, 211613-69, 211613-7, 211613-70, 211613-71, 211613-72, 211613-73, 211613-74, 211613-75, 211613-76, 211613-77, 211613-78, 211613-79, 211613-8, 211613-80, 211613-81, 211613-82, 211613-83, 211613-84, 211613-85, 211613-86, 211613-87, 211613-88, 211613-89, 211613-9, 211613-90, 211613-91, 211613-92, 211613-93, 211613-94, 211613-95, 211613-96, 211613-97, 211613-98, 211613-99, , 211577-1, 211577-10, 211577-11, 211577-12, 211577-13, 211577-14, 211577-15, 211577-16, 211577-17, 211577-18, 211577-19, 211577-2, 211577-20, 211577-21, 211577-22, 211577-23, 211577-24, 211577-25, 211577-26, 211577-27, 211577-28, 211577-29, 211577-3, 211577-30, 211577-31, 211577-32, 211577-33, 211577-34, 211577-35, 211577-36, 211577-37, 211577-38, 211577-39, 211577-4, 211577-40, 211577-41, 211577-42, 211577-43, 211577-44, 211577-45, 211577-46, 211577-47, 211577-48, 211577-49, 211577-5, 211577-50, 211577-51, 211577-52, 211577-53, 211577-54, 211577-55, 211577-56, 211577-57, 211577-58, 211577-59, 211577-6, 211577-60, 211577-61, 211577-62, 211577-63, 211577-64, 211577-65, 211577-66, 211577-67, 211577-68, 211577-69, 211577-7, 211577-70, 211577-71, 211577-72, 211577-73, 211577-74, 211577-75, 211577-76, 211577-77, 211577-78, 211577-79, 211577-8, 211577-80, 211577-81, 211577-82, 211577-83, 211577-84, 211577-85, 211577-86, 211577-87, 211577-88, 211577-89, 211577-9, 211577-90, 211577-91, 211577-92, 211577-93, 211577-94, 211577-95, 211577-96, 211577-97, 211577-98, 211577-99, , 211576-1, 211576-10, 211576-11, 211576-12, 211576-13, 211576-14, 211576-15, 211576-16, 211576-17, 211576-18, 211576-19, 211576-2, 211576-20, 211576-23, 211576-24, 211576-25, 211576-26, 211576-27, 211576-28, 211576-29, 211576-30, 211576-31, 211576-33, 211576-34, 211576-35, 211576-36, 211576-37, 211576-38, 211576-39, 211576-41, 211576-42, 211576-43, 211576-44, 211576-45, 211576-46, 211576-47, 211576-48, 211576-49, 211576-5, 211576-51, 211576-52, 211576-53, 211576-54, 211576-55, 211576-58, 211576-59, 211576-6, 211576-60, 211576-61, 211576-63, 211576-64, 211576-66, 211576-67, 211576-68, 211576-69, 211576-7, 211576-70, 211576-71, 211576-76, 211576-78, 211576-79, 211576-8, 211576-80, 211576-81, 211576-82, 211576-83, 211576-84, 211576-85, 211576-86, 211576-87, 211576-89, 211576-9, 211576-90, 211576-91, 211576-92, 211576-94, , 2B10865-1, 2B10865-2, 2B10865-3, 2B10865-4, 2B10865-5, 2B10865-6, 2B10865-7, 2B10865-8, 2B10865-9, 2B10865-10, 2B10865-11, 2B10865-12, 2B10865-13, 2B10865-14, 2B10865-15, 2B10865-16, 2B10865-17, 2B10865-18, 2B10865-19, 2B10865-20, 2B10865-21, 2B10865-22, 2B10865-23, 2B10865-24, 2B10865-25, 2B10865-26, 2B10865-27, 2B10865-28, 2B10865-29, 2B10865-30, 2B10865-31, 2B10865-32, 2B10865-33, 2B10865-34, 2B10865-35, 2B10865-36, 2B10865-37, 2B10865-38, 2B10865-39, 2B10865-40, 2B10865-41, 2B10865-42, 2B10865-43, 2B10865-44, 2B10865-45, 2B10865-46, 2B10865-47, 2B10865-48, 2B10865-49, 2B10865-50, 2B10865-51, 2B10865-52, 2B10865-53, 2B10865-54, 2B10865-55, 2B10865-56, 2B10865-57, 2B10865-58, 2B10865-59, 2B10865-60, 2B10865-61, 2B10865-62, 2B10865-63, 2B10865-64, 2B10865-65, 2B10865-66, 2B10865-67, 2B10865-68, 2B10865-69, 2B10865-70, 2B10865-71, 2B10865-72, 2B10865-73, 2B10865-74, 2B10865-75, 2B10865-76, 2B10865-77, 2B10865-78, 2B10865-79, 2B10865-80, 2B10865-81, 2B10865-82, 2B10865-83, 2B10865-84, 2B10865-85, 2B10865-86, 2B10865-87, 2B10865-88, 2B10865-89, 2B10865-90, 2B10865-91, 2B10865-92, 2B10865-93, 2B10865-94, 2B10865-95, 2B10865-96, 2B10865-97, 2B10865-98, 2B10865-99, 2B10865-100, 2B10865-101, 2B10865-102, 2B10865-103, 2B10865-104, 2B10865-105, 2B10865-106, 2B10865-107, 2B10865-108, 2B10865-109, 2B10865-110, 2B10865-111, 2B10865-112, 2B10865-113, 2B10865-114, 2B10865-115, 2B10865-116, 2B10865-117, 2B10865-118, 2B10865-119, 2B10865-120, , 2B10864-1, 2B10864-2, 2B10864-3, 2B10864-4, 2B10864-5, 2B10864-6, 2B10864-7, 2B10864-8, 2B10864-9, 2B10864-10, 2B10864-11, 2B10864-12, 2B10864-13, 2B10864-14, 2B10864-15, 2B10864-16, 2B10864-17, 2B10864-18, 2B10864-19, 2B10864-20, 2B10864-21, 2B10864-22, 2B10864-23, 2B10864-24, 2B10864-25, 2B10864-26, 2B10864-27, 2B10864-28, 2B10864-29, 2B10864-30, 2B10864-31, 2B10864-32, 2B10864-33, 2B10864-34, 2B10864-35, 2B10864-36, 2B10864-37, 2B10864-38, 2B10864-39, 2B10864-40, 2B10864-41, 2B10864-42, 2B10864-43, 2B10864-44, 2B10864-45, 2B10864-46, 2B10864-47, 2B10864-48, 2B10864-49, 2B10864-50, 2B10864-51, 2B10864-52, 2B10864-53, 2B10864-54, 2B10864-55, 2B10864-56, 2B10864-57, 2B10864-58, 2B10864-59, 2B10864-60, 2B10864-61, 2B10864-62, 2B10864-63, 2B10864-64, 2B10864-65, 2B10864-66, 2B10864-67, 2B10864-68, 2B10864-69, 2B10864-70, 2B10864-71, 2B10864-72, 2B10864-73, 2B10864-74, 2B10864-75, 2B10864-76, 2B10864-77, 2B10864-78, 2B10864-79, 2B10864-80, 2B10864-81, 2B10864-82, 2B10864-83, 2B10864-84, 2B10864-85, 2B10864-86, 2B10864-87, 2B10864-88, 2B10864-89, 2B10864-90, 2B10864-91, 2B10864-92, 2B10864-93, 2B10864-94, 2B10864-95, 2B10864-96, 2B10864-97, 2B10864-98, 2B10864-99, 2B10864-100, 2B10864-101, 2B10864-102, 2B10864-103, 2B10864-104, 2B10864-105, 2B10864-106, 2B10864-107, 2B10864-108, 2B10864-109, 2B10864-110, 2B10864-111, 2B10864-112, , 210859-1, 210859-10, 210859-11, 210859-12, 210859-13, 210859-14, 210859-15, 210859-16, 210859-17, 210859-18, 210859-19, 210859-2, 210859-20, 210859-21, 210859-22, 210859-23, 210859-24, 210859-25, 210859-26, 210859-27, 210859-28, 210859-29, 210859-3, 210859-30, 210859-31, 210859-32, 210859-33, 210859-34, 210859-35, 210859-36, 210859-37, 210859-38, 210859-39, 210859-4, 210859-40, 210859-41, 210859-42, 210859-43, 210859-44, 210859-45, 210859-46, 210859-47, 210859-48, 210859-49, 210859-5, 210859-50, 210859-51, 210859-52, 210859-53, 210859-54, 210859-55, 210859-56, 210859-57, 210859-58, 210859-59, 210859-6, 210859-60, 210859-61, 210859-62, 210859-63, 210859-64, 210859-65, 210859-66, 210859-67, 210859-68, 210859-69, 210859-7, 210859-70, 210859-71, 210859-72, 210859-73, 210859-74, 210859-75, 210859-76, 210859-77, 210859-78, 210859-79, 210859-8, 210859-80, 210859-81, 210859-82, 210859-83, 210859-84, 210859-9, , 210803-1, 210803-10, 210803-11, 210803-12, 210803-13, 210803-14, 210803-15, 210803-16, 210803-17, 210803-18, 210803-19, 210803-2, 210803-20, 210803-21, 210803-22, 210803-23, 210803-24, 210803-25, 210803-26, 210803-27, 210803-28, 210803-29, 210803-3, 210803-30, 210803-31, 210803-32, 210803-33, 210803-34, 210803-35, 210803-36, 210803-37, 210803-38, 210803-39, 210803-4, 210803-40, 210803-41, 210803-42, 210803-43, 210803-44, 210803-45, 210803-46, 210803-47, 210803-48, 210803-49, 210803-5, 210803-50, 210803-51, 210803-52, 210803-53, 210803-54, 210803-55, 210803-56, 210803-57, 210803-58, 210803-59, 210803-6, 210803-60, 210803-61, 210803-7, 210803-8, 210803-9, , 210436-1, 210436-10, 210436-11, 210436-12, 210436-13, 210436-14, 210436-15, 210436-16, 210436-17, 210436-18, 210436-19, 210436-2, 210436-20, 210436-21, 210436-22, 210436-23, 210436-24, 210436-25, 210436-26, 210436-27, 210436-28, 210436-29, 210436-3, 210436-30, 210436-31, 210436-32, 210436-33, 210436-34, 210436-35, 210436-36, 210436-37, 210436-38, 210436-39, 210436-4, 210436-40, 210436-41, 210436-42, 210436-43, 210436-44, 210436-45, 210436-46, 210436-47, 210436-48, 210436-49, 210436-5, 210436-50, 210436-51, 210436-52, 210436-53, 210436-54, 210436-55, 210436-56, 210436-57, 210436-58, 210436-59, 210436-6, 210436-60, 210436-61, 210436-62, 210436-63, 210436-64, 210436-65, 210436-66, 210436-67, 210436-7, 210436-8, 210436-9, , 210435-1, 210435-10, 210435-11, 210435-12, 210435-13, 210435-14, 210435-15, 210435-16, 210435-17, 210435-18, 210435-19, 210435-2, 210435-20, 210435-21, 210435-22, 210435-23, 210435-24, 210435-25, 210435-26, 210435-27, 210435-28, 210435-29, 210435-3, 210435-30, 210435-31, 210435-32, 210435-33, 210435-34, 210435-35, 210435-36, 210435-37, 210435-38, 210435-39, 210435-4, 210435-40, 210435-41, 210435-42, 210435-43, 210435-44, 210435-45, 210435-46, 210435-47, 210435-48, 210435-49, 210435-5, 210435-50, 210435-51, 210435-52, 210435-53, 210435-54, 210435-55, 210435-56, 210435-57, 210435-58, 210435-59, 210435-6, 210435-60, 210435-61, 210435-62, 210435-63, 210435-64, 210435-65, 210435-66, 210435-67, 210435-68, 210435-69, 210435-7, 210435-70, 210435-71, 210435-72, 210435-73, 210435-74, 210435-75, 210435-76, 210435-77, 210435-78, 210435-79, 210435-8, 210435-80, 210435-81, 210435-82, 210435-83, 210435-84, 210435-85, 210435-86, 210435-87, 210435-88, 210435-89, 210435-9, 210435-90, 210435-91, 210435-92, , 210428-1, 210428-10, 210428-11, 210428-12, 210428-13, 210428-14, 210428-15, 210428-16, 210428-17, 210428-18, 210428-19, 210428-2, 210428-20, 210428-21, 210428-22, 210428-23, 210428-24, 210428-25, 210428-26, 210428-27, 210428-28, 210428-29, 210428-3, 210428-30, 210428-31, 210428-32, 210428-33, 210428-34, 210428-35, 210428-36, 210428-37, 210428-38, 210428-39, 210428-4, 210428-40, 210428-41, 210428-42, 210428-43, 210428-44, 210428-5, 210428-6, 210428-7, 210428-8, 210428-9, , 210426-1, 210426-10, 210426-11, 210426-12, 210426-13, 210426-14, 210426-15, 210426-16, 210426-17, 210426-18, 210426-19, 210426-2, 210426-20, 210426-21, 210426-22, 210426-23, 210426-24, 210426-25, 210426-26, 210426-27, 210426-28, 210426-29, 210426-3, 210426-30, 210426-31, 210426-32, 210426-33, 210426-34, 210426-35, 210426-36, 210426-37, 210426-38, 210426-39, 210426-4, 210426-40, 210426-41, 210426-42, 210426-43, 210426-44, 210426-45, 210426-46, 210426-47, 210426-48, 210426-49, 210426-5, 210426-50, 210426-51, 210426-52, 210426-53, 210426-54, 210426-55, 210426-56, 210426-57, 210426-58, 210426-59, 210426-6, 210426-60, 210426-61, 210426-62, 210426-63, 210426-64, 210426-65, 210426-66, 210426-67, 210426-68, 210426-69, 210426-7, 210426-70, 210426-71, 210426-72, 210426-73, 210426-74, 210426-75, 210426-76, 210426-77, 210426-78, 210426-79, 210426-8, 210426-80, 210426-81, 210426-82, 210426-83, 210426-84, 210426-85, 210426-86, 210426-87, 210426-9, 210426-94, 210426-95, 210426-96, 210426-97, 210426-98, , 210425-20, 210425-21, 210425-30, 210425-32, 210425-4, 210425-45, 210425-5, 210425-50, 210425-58, 210425-60, 210425-65, 210425-66, 210425-71, 210425-72, 210425-75, 210425-78, 210425-86, , 209705-1, 209705-2, 209705-3, 209705-4, 209705-5, , 209703-12, 209703-14, 209703-15, 209703-16, , 209556-1, 209556-10, 209556-11, 209556-12, 209556-13, 209556-14, 209556-15, 209556-16, 209556-17, 209556-18, 209556-19, 209556-2, 209556-20, 209556-21, 209556-22, 209556-23, 209556-24, 209556-25, 209556-26, 209556-27, 209556-28, 209556-29, 209556-3, 209556-30, 209556-31, 209556-32, 209556-33, 209556-34, 209556-35, 209556-36, 209556-37, 209556-38, 209556-39, 209556-4, 209556-40, 209556-41, 209556-42, 209556-43, 209556-44, 209556-45, 209556-5, 209556-6, 209556-7, 209556-8, 209556-9, , 209555-23, 209555-24, 209555-25, 209555-26, 209555-27, 209555-28, 209555-29, 209555-30, 209555-31, 209555-32, 209555-33, 209555-34, 209555-35, 209555-36, 209555-37, 209555-38, 209555-39, 209555-40, 209555-41, 209555-42, 209555-43, 209555-44, 209555-6, 209555-8, , 201245-1, 201245-10, 201245-11, 201245-12, 201245-13, 201245-14, 201245-15, 201245-16, 201245-17, 201245-18, 201245-19, 201245-2, 201245-20, 201245-21, 201245-22, 201245-3, 201245-4, 201245-5, 201245-6, 201245-7, 201245-8, 201245-9, , 201244-1, 201244-10, 201244-11, 201244-12, 201244-13, 201244-14, 201244-15, 201244-16, 201244-17, 201244-18, 201244-19, 201244-2, 201244-20, 201244-21, 201244-22, 201244-23, 201244-24, 201244-25, 201244-26, 201244-27, 201244-28, 201244-29, 201244-3, 201244-30, 201244-31, 201244-32, 201244-33, 201244-34, 201244-35, 201244-36, 201244-37, 201244-38, 201244-39, 201244-4, 201244-40, 201244-41, 201244-42, 201244-43, 201244-44, 201244-45, 201244-46, 201244-47, 201244-48, 201244-49, 201244-5, 201244-50, 201244-6, 201244-7, 201244-8, 201244-9, , 201243-11, 201243-12, 201243-14, 201243-19, 201243-20, 201243-21, 201243-23, 201243-30, 201243-33, 201243-52, 201243-60, 201243-61; EXPANSION - additional serial numbers included:, 2B11665-67, 2B11665-68, 2B11664-72, 210425-69, 210425-70, 2B11664-48, 2B11664-75, 2B11664-76, 2B11664-88, 2B11665-12, 2B11665-29, 2B11665-32, 2B11665-34, 2B11665-35, 2B11665-36, 2B11665-7, 2B11807-22, 2B11807-23, 2B11808-69, 2B11808-7, 2B11808-71, 2B11808-72, 2B11808-70, 2B11664-71, 210425-41, 2B11808-13, 2B11664-87, 211576-62, 2B11664-33, 2B11665-78, 2B11665-81, 2B11665-84, 2B11665-85, 2B11665-9, 2B11807-42, 2B11807-43, 2B11808-17, 2B11808-53, 2B11876-64, 2B11808-12, 210425-68, 2B11808-86, 2B11808-49, 211576-40, 211576-50, 211576-56, 2B11876-41, 2B11876-31, 210425-52, 211576-74, 2B11664-1, 2B11665-80, 2B11808-80, 2B11665-13, 2B11807-12, 2B11807-8, 2B11876-35, 2B11876-36, 2B11876-37, 2B11876-38, 2B11876-43, 2B11664-3, 2B11665-64, 2B11807-3, 2B11808-27, 2B11808-28, 2B11808-3, 2B11664-19, 2B11808-47, 211576-75, 211576-88, 211576-93, 211576-95, 2B11664-60, 2B11808-1, 2B11808-10, 2B11808-20, 2B11808-30, 2B11808-4, 2B11808-40, 210425-28, 2B11664-77, 2B11665-50, 2B11665-54, 2B11876-34, 2B11664-74, 2B11664-47, 2B11664-51, 210425-83, 2B11878-88, 2B11876-82, 2B11876-83, 2B11664-22, 2B11665-65, 2B11807-28, 2B11807-29, 2B11807-31, 2B11808-26, 2B11878-5, 2B11876-50, 2B11807-15, 2B11808-51, 2B11808-52, 2B11665-74, 2B11876-11, 2B11876-12, 210425-49, 210425-44, 2B11665-42, 2B11807-41, 2B11665-48, 2B11807-14, 2B11878-87, 2B11808-54, 2B11665-22, 2B11665-27, 2B11665-83, 2B11808-18, 2B11664-55, 210425-40, 210425-87, 2B11664-62, 210425-43, 2B11876-73, 2B11808-87, 2B11808-9, 2B11876-84, 2B11876-85, 2B11876-86, 2B11665-61, 2B11876-48, 210425-85, 2B11665-71, 2B11876-25, 2B11876-26, 2B11665-17, 2B11665-19, 2B11808-32, 2B11808-36, 2B11876-28, 2B11876-29, 2B11876-60, 2B11665-33, 211576-73, 2B11664-32, 2B11876-40, 2B11664-50, 2B11664-53, 2B11876-24, 2B11878-4, 2B11876-75, 2B11876-69, 210425-48, 2B11876-53, 210425-42, 2B11664-70, 2B11876-4, 2B11664-40, 210425-46, 210425-47, 210425-81, 2B11808-15, 2B11808-78, 210425-8, 2B11807-9, 2B11808-19, 2B11808-2, 2B11808-22, 2B11664-10, 2B11665-2, 2B11665-40, 2B11665-82, 2B11807-38, 2B11807-39, 2B11808-5, 2B11808-58, 2B11876-54, 2B11876-55, 2B11876-58, 2B11876-6, 210425-26, 2B11664-43, 2B11665-11, 2B11664-92, 2B11807-13, 2B11807-16, 2B11807-17, 2B11807-27, 2B11664-79, 2B11664-82, 2B11665-5, 2B11665-55, 2B11665-56, 2B11665-57, 2B11808-23, 2B11876-16, 2B11876-30, 2B11876-32, 2B11876-33, 210425-27, 210425-53, 210425-54, 210425-55, 2B11665-38, 2B11665-39, 2B11665-53, 2B11876-67, 2B11876-78, 211576-72, 2B11665-59, 2B11665-63, 211576-32, 2B11878-6, 210425-80, 2B11808-46, 2B11808-73, 2B11808-76, 2B11876-5, 2B11876-59, 2B11878-7, 2B11878-8, 2B11807-19, 2B11876-13, 2B11876-15, 2B11876-68, 2B11876-7, 2B11876-81, 210425-67, 2B11665-45, 2B11665-52, 2B11807-30, 2B11808-59, 210425-56, 210425-57, 2B11664-18, 2B11808-45, 2B11807-6, 2B11665-30, 2B11665-31, 2B11665-79, 2B11808-41, 2B11664-67, 210425-51, 2B11665-46, 210425-25, 2B11807-2, 2B11807-21, 2B11876-70, 2B11665-4, 2B11808-55, 2B11664-11, 2B11807-4, 2B11808-56, 2B11876-74, 2B11808-29, 2B11808-11, 211576-21, 2B11808-31, 2B11807-18, 211576-96, 2B11878-9, 2B11876-87, 2B11876-91, 2B11665-1, 2B11878-85, 2B11664-9, 210425-59, 2B11665-18, 2B11808-14, 2B11808-33, 2B11808-34, 2B11808-37, 2B11808-38, 2B11807-40, 210425-84, 2B11807-32, 2B11807-34, 210425-9, 2B11665-3, 2B11876-1, 2B11876-10, 2B11665-47, 2B11808-74, 2B11807-11, 2B11808-75, 2B11878-1, 210425-29, 2B11665-10, 2B11808-35, 2B11808-39, 2B11808-16, 210425-35, 210425-36, 211576-22, 211576-57, 2B11664-44, 2B11665-49, 2B11664-52, 2B11807-24, 2B11665-51, 2B11876-21, 2B11876-76, 2B11876-8, 2B11664-90, 2B11665-72, 2B11665-76, 2B11665-77, 210425-37, 210425-39, 2B11664-30, 2B11808-50, 2B11808-68, 2B11876-93, 2B11664-16, 2B11808-6, 2B11664-20, 2B11876-92, 2B11664-57, 2B11664-61, 2B11664-58, 2B11807-20, 2B11665-58, 2B11665-62, Expansion additional serial numbers added: 2B11876-54, 2B11876-55, 2B11876-58, 2B11876-59, 2B11876-6, 2B11876-60, 2B11876-64, 2B11876-67, 2B11876-68, 2B11876-69, 2B11876-7, 2B11876-70, 2B11876-73, 2B11876-74, 2B11876-75, 2B11876-76, 2B11876-78, 2B11876-8, 2B11876-81, 2B11876-82, 2B11876-83, 2B11876-84, 2B11876-85, 2B11876-86, 2B11876-87, 2B11876-91, 2B11876-92, 2B11876-93, 2B11878-1, 2B11878-4, 2B11878-5, 2B11878-6, 2B11878-7, 2B11878-8, 2B11878-85, 2B11878-87, 2B11878-88, 2B11878-9, 211569-46, 211569-6, 211569-64, 2B11768-1, 2B11768-11, 2B11768-12, 2B11768-13, 2B11768-14, 2B11768-15, 2B11768-16, 2B11768-17, 2B11768-18, 2B11768-19, 2B11768-2, 2B11768-20, 2B11768-21, 2B11768-22, 2B11768-23, 2B11768-24, 2B11768-25, 2B11768-27, 2B11768-28, 2B11768-29, 2B11768-3, 2B11768-30, 2B11768-31, 2B11768-32, 2B11768-33, 2B11768-35, 2B11768-36, 2B11768-38, 2B11768-39, 2B11768-4, 2B11768-40, 2B11768-41, 2B11768-42, 2B11768-45, 2B11768-47, 2B11768-51, 2B11768-52, 2B11768-56, 2B11768-57, 2B11768-58, 2B11768-59, 2B11768-6, 2B11768-60, 2B11768-61, 2B11768-62, 2B11768-63, 2B11768-64, 2B11768-65, 2B11768-66, 2B11768-67, 2B11768-68, 2B11768-69, 2B11768-7, 2B11768-70, 2B11768-72, 2B11768-74, 2B11768-75, 2B11768-76, 2B11768-77, 2B11768-78, 2B11768-79, 2B11768-8, 2B11768-80, 2B11768-81, 2B11768-83, 2B11768-84, 2B11768-85, 2B11768-86, 2B11768-87, 2B11768-88, 2B11768-89, 2B11768-9, 2B11768-92, 2B11768-93, 2B11768-94, 2B11768-95, 2B11768-96, 2B11806-10, 2B11806-11, 2B11806-12, 2B11806-13, 2B11806-14, 2B11806-17, 2B11806-19, 2B11806-20, 2B11806-22, 2B11806-23, 2B11806-24, 2B11806-25, 2B11806-26, 2B11806-27, 2B11806-29, 2B11806-3, 2B11806-30, 2B11806-31, 2B11806-32, 2B11806-33, 2B11806-34, 2B11806-39, 2B11806-4, 2B11806-40, 2B11806-48, 2B11806-49, 2B11806-5, 2B11806-51, 2B11806-53, 2B11806-54, 2B11806-55, 2B11806-58, 2B11806-59, 2B11806-6, 2B11806-61, 2B11806-62, 2B11806-63, 2B11806-64, 2B11806-65, 2B11806-66, 2B11806-67, 2B11806-68, 2B11806-69, 2B11806-7, 2B11806-70, 2B11806-71, 2B11806-72, 2B11806-74, 2B11806-75, 2B11806-76, 2B11806-79, 2B11806-80, 2B11806-81, 2B11806-82, 2B11806-83, 2B11806-84, 2B11806-85, 2B11806-86, 2B11806-87, 2B11806-88, 2B11806-89, 2B11806-9, 2B11806-91, 2B11806-94, 2B11806-95, 2B11806-96, 2B11806-97, 2B11879-1, 2B11879-10, 2B11879-11, 2B11879-14, 2B11879-15, 2B11879-16, 2B11879-18, 2B11879-19, 2B11879-2, 2B11879-20, 2B11879-21, 2B11879-22, 2B11879-23, 2B11879-24, 2B11879-27, 2B11879-28, 2B11879-29, 2B11879-3, 2B11879-30, 2B11879-31, 2B11879-34, 2B11879-35, 2B11879-36, 2B11879-37, 2B11879-38, 2B11879-39, 2B11879-4, 2B11879-40, 2B11879-41, 2B11879-42, 2B11879-43, 2B11879-44, 2B11879-45, 2B11879-46, 2B11879-47, 2B11879-49, 2B11879-51, 2B11879-53, 2B11879-54, 2B11879-55, 2B11879-56, 2B11879-57, 2B11879-58, 2B11879-59, 2B11879-61, 2B11879-62, 2B11879-63, 2B11879-64, 2B11879-65, 2B11879-66, 2B11879-67, 2B11879-69, 2B11879-7, 2B11879-70, 2B11879-71, 2B11879-72, 2B11879-73, 2B11879-74, 2B11879-75, 2B11879-76, 2B11879-77, 2B11879-80, 2B11879-81, 2B11879-82, 2B11879-83, 2B11879-85, 2B11879-86, 2B11879-87, 2B11879-88, 2B11879-89, 2B11879-9, 2B11879-90, 2B11879-91, 2B11879-92, 2B11879-93, 2B11879-94, 2B11879-95, 2B11880-1, 2B11880-10, 2B11880-11, 2B11880-12, 2B11880-13, 2B11880-14, 2B11880-15, 2B11880-16, 2B11880-17, 2B11880-18, 2B11880-19, 2B11880-2, 2B11880-22, 2B11880-23, 2B11880-25, 2B11880-26, 2B11880-27, 2B11880-28, 2B11880-3, 2B11880-30, 2B11880-32, 2B11880-34, 2B11880-35, 2B11880-39, 2B11880-4, 2B11880-40, 2B11880-41, 2B11880-42, 2B11880-44, 2B11880-46, 2B11880-47, 2B11880-48, 2B11880-49, 2B11880-5, 2B11880-52, 2B11880-55, 2B11880-59, 2B11880-6, 2B11880-7, 2B11880-8, 2B11880-9.
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KY, KS, LA, MD, MO, MI, MN, MA, MS, NC, NM, ND, NV, NY, NE, NJ, OK, OH, OR, PR, PA, RI, SC, TN, TX, UT, VA, WA, WI and the countries of The Netherlands, Belgium, Slovakia, Switzerland, Germany, United Kingdom, Argentina, Italy, Spain, South Africa, Iran, Israel, and Czech Republic.
1550 units