FDA Enforcement Class I Ongoing

Alaris System PC Unit Model 8015

Recall: Z-2909-2020 · Reported September 16, 2020

Enforcement

Recall Number
Z-2909-2020
Event ID
85900
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
CareFusion 303, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 16, 2020
Initiation Date
August 4, 2020
Classification Date
September 4, 2020
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386, United States

Description

Alaris System PC Unit Model 8015

Reason

The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.

Code Info

All serial numbers manufactured from April 07, 2017 to Present.

Distribution

U.S.:LA, FL, KY, MI, AL, GA, CA, TN, NC, TX, MA, OH, CO, WI, IL, KS, MO, UT, AK, MD, NY, NV, VA, NE, IN, PA, VT, DC, SC, CT, AR, NJ, MT, ND ,HI , NH, WA, DE, ME, OR, MN, AZ, MS, IA, NM, WY, ID, OK, SD, WV, MP, GU O.U.S.: CA, AE, AU, BE, CA, CH, IL, IN, KW, N/A, NL, NZ, PH, QA, SA, SG, TW, US, ZA

Quantity

313740