Nerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMG ET Tube, 6.0mm REF NVTKIT-C6-5, Rx Only, Sterile, UDI: B006NVTKITC652/$$7061918A - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
Enforcement
- Recall Number
- Z-2900-2020
- Event ID
- 86102
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Neurovision Medical Products Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 2, 2020
- Initiation Date
- July 21, 2020
- Classification Date
- August 27, 2020
- Termination Date
- April 30, 2021
- Address
- 2443 Portola Rd, N/A, Ventura, CA, 93003-7726, United States
Description
Nerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMG ET Tube, 6.0mm REF NVTKIT-C6-5, Rx Only, Sterile, UDI: B006NVTKITC652/$$7061918A - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
Kit Lots 070919A 072619A, 073119A 082019B, 120219D 120319G 120519A 121819E 121919E 122019B 010820D 011520A 011620A 012320D 012420B 030620A 031220E 052120B 062620A 022820A 041620C 062520A
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, MI, MN, MO, MS, NE, NY, OH, OR, PA, SC, SD, TN, TX, WA, WI and the countries of Canada, Dominican Republic, Hong Kong.
1,089