FDA Enforcement Class II Ongoing

Philips SureSigns VS2 monitor, 863079 863080 863081 863082 863283 863286 The Sure Signs VS2 vital signs monitor is for use by healthcare professionals whenever there is a need for monitoring the physiological parameter of patients. Standard and optional parameters include: NBP, Sp02, Temperature. The Sure Signs VS2 vital signs monitor is for the monitoring, recording and alarming of multiple parameters of adult, pediatric and neonates in healthcare environments. Additionally, this monitor is intended for transport situations within the healthcare facility.

Recall: Z-2896-2018 · Reported August 29, 2018

Enforcement

Recall Number
Z-2896-2018
Event ID
80645
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips Electronics North America Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 29, 2018
Initiation Date
July 13, 2018
Classification Date
August 22, 2018
Address
3000 Minuteman Rd, Andover, MA, 01810-1032, United States

Description

Philips SureSigns VS2 monitor, 863079 863080 863081 863082 863283 863286 The Sure Signs VS2 vital signs monitor is for use by healthcare professionals whenever there is a need for monitoring the physiological parameter of patients. Standard and optional parameters include: NBP, Sp02, Temperature. The Sure Signs VS2 vital signs monitor is for the monitoring, recording and alarming of multiple parameters of adult, pediatric and neonates in healthcare environments. Additionally, this monitor is intended for transport situations within the healthcare facility.

Reason

Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.

Code Info

UDI# VS2 (863079): 00884838009639, (863080): 00884838009646, (863081): 00884838009653, (863082): 00884838009660

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Cambodia, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Czech Republic, Dominican Republic, El Salvador, Egypt, Estonia, Ethiopia, France, French Guiana, French Polynesia, Finland, Guadeloupe, Germany, Ghana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Kenya, Korea, Kosovo, Republic of, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Lybia, Macao, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, Zambia and Zimbabwe.

Quantity

VM 96,324 Units; VS 110,424 Units; VSV 691 Units