FDA Enforcement Class II Ongoing

Flexible Intubation Fiberscope, Part: 11301BN1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.

Recall: Z-2892-2020 · Reported September 2, 2020

Enforcement

Recall Number
Z-2892-2020
Event ID
86009
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Karl Storz Endoscopy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 2, 2020
Initiation Date
July 27, 2020
Classification Date
August 27, 2020
Address
2151 E Grand Ave, N/A, El Segundo, CA, 90245-5017, United States

Description

Flexible Intubation Fiberscope, Part: 11301BN1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.

Reason

Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.

Code Info

Instruction Manual Version: 08/2018

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.

Quantity

1246