Philips SureSigns VSV (Vital Signs Viewer), 863067 Indicated for central monitoring of multiple adult, pediatric and neonatal patients, and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric and neonatal patients to gain information for the treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.
Enforcement
- Recall Number
- Z-2892-2018
- Event ID
- 80645
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Philips Electronics North America Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 29, 2018
- Initiation Date
- July 13, 2018
- Classification Date
- August 22, 2018
- Address
- 3000 Minuteman Rd, Andover, MA, 01810-1032, United States
Description
Philips SureSigns VSV (Vital Signs Viewer), 863067 Indicated for central monitoring of multiple adult, pediatric and neonatal patients, and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric and neonatal patients to gain information for the treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
UDI# VSV (863067): 00884838009578
Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Cambodia, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Czech Republic, Dominican Republic, El Salvador, Egypt, Estonia, Ethiopia, France, French Guiana, French Polynesia, Finland, Guadeloupe, Germany, Ghana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Kenya, Korea, Kosovo, Republic of, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Lybia, Macao, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, Zambia and Zimbabwe.
VM 96,324 Units; VS 110,424 Units; VSV 691 Units