Flexible Intubation Fiberscope, Part: 11301AA1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Enforcement
- Recall Number
- Z-2890-2020
- Event ID
- 86009
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Karl Storz Endoscopy
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 2, 2020
- Initiation Date
- July 27, 2020
- Classification Date
- August 27, 2020
- Address
- 2151 E Grand Ave, N/A, El Segundo, CA, 90245-5017, United States
Description
Flexible Intubation Fiberscope, Part: 11301AA1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
Instruction Manual Version: 08/2018
Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.