Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.
Recall: Z-2888-2016
·
Reported October 5, 2016
Enforcement
- Recall Number
- Z-2888-2016
- Event ID
- 75160
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Spiegelberg Gmbh & Co. KG
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 5, 2016
- Initiation Date
- December 29, 2009
- Classification Date
- September 26, 2016
- Termination Date
- October 5, 2016
- Address
- Tempowerkring 4, N/A, Hamburg, N/A, N/A, Germany
Description
Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.
Reason
In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.
Code Info
Probe 3PN SND1 3.1 .53/FV534P: SNs: 12835-1 3393; Probe 3PS SND13.1 .63/FV535P: SNs: 6937-7122
Distribution
Nationwide Distribution to PA only
Quantity
22 units