FDA Enforcement
Class II
Terminated
VITROS XT3400 Chemistry System, Product code 6844458
Recall: Z-2878-2020
·
Reported September 2, 2020
Enforcement
- Recall Number
- Z-2878-2020
- Event ID
- 86096
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ortho Clinical Diagnostics
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 2, 2020
- Initiation Date
- July 20, 2020
- Classification Date
- August 24, 2020
- Termination Date
- December 18, 2023
- Address
- 100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States
Description
VITROS XT3400 Chemistry System, Product code 6844458
Reason
When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous results.
Code Info
Serial numbers J34500122 J34500231
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Quantity
pending