FDA Enforcement Class II Terminated

Connecting Bolt, Model 14-442093

Recall: Z-2875-2018 · Reported August 29, 2018

Enforcement

Recall Number
Z-2875-2018
Event ID
80543
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer Biomet, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 29, 2018
Initiation Date
June 18, 2018
Classification Date
August 17, 2018
Termination Date
March 26, 2020
Address
56 E Bell Dr, Warsaw, IN, 46582-6989, United States

Description

Connecting Bolt, Model 14-442093

Reason

Product was manufactured from the incorrect material, which shows a potential decrease in the fatigue strength of the connecting bolt that can lead to fracture.

Code Info

...

Distribution

AR, AZ, CA, FL, GA, IN, KY, LA, MA, MI, MN, MO, MT, NJ, NY, OH, SC, TX, UT & WI

Quantity

247 units