FDA Enforcement Class II Terminated

VITROS Chemistry Products BuBc Slides - in vitro diagnostic use only. VITROS Chemistry Products BuBc Slides quantitatively measure unconjugated bilirubin (Bu) and conjugated bilirubin (Bc) concentrations in serum and plasma, Product code: 8383051 (60 slide); 1612365 (18 slide)

Recall: Z-2874-2020 · Reported September 2, 2020

Enforcement

Recall Number
Z-2874-2020
Event ID
86070
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 2, 2020
Initiation Date
July 13, 2020
Classification Date
August 24, 2020
Termination Date
July 27, 2021
Address
513 Technology Blvd, N/A, Rochester, NY, 14626-3601, United States

Description

VITROS Chemistry Products BuBc Slides - in vitro diagnostic use only. VITROS Chemistry Products BuBc Slides quantitatively measure unconjugated bilirubin (Bu) and conjugated bilirubin (Bc) concentrations in serum and plasma, Product code: 8383051 (60 slide); 1612365 (18 slide)

Reason

Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides

Code Info

All Current, Future and Past Lots UDI: 10758750004829, 10758750009398

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Bermuda HM08, Brazil, Canada L3R 4G5, China, Denmark, France, Germany, Italy, Mexico, Norway, Poland, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.

Quantity

80351:US=35113; OUS: OUS=1593941.1612365: US=20068; OUS= 35534