FDA Enforcement Class II Terminated

BD BBL GC-Lect Agar; Cat. No. 297715 The device is a selective medium providing enhanced growth and recovery of Neisseria gonorrhoeae and better inhibition of contamination bacteria and fungi.

Recall: Z-2869-2018 · Reported August 29, 2018

Enforcement

Recall Number
Z-2869-2018
Event ID
80573
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Becton Dickinson & Co.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 29, 2018
Initiation Date
June 21, 2018
Classification Date
August 17, 2018
Termination Date
July 9, 2019
Address
7 Loveton Cir, BD Diagnostic Systems, Sparks, MD, 21152-9212, United States

Description

BD BBL GC-Lect Agar; Cat. No. 297715 The device is a selective medium providing enhanced growth and recovery of Neisseria gonorrhoeae and better inhibition of contamination bacteria and fungi.

Reason

A portion of this lot was manufactured using Gentamicin instead of Vancomycin. Gentamicin is an inhibitory agent of Neisseria organisms. This product is a selective plated medium to provide enhanced growth and recovery of N gonorrhoeae. Inhibited growth may cause a false negative result or a delayed result leading to incorrect or delayed treatment.

Code Info

8110781

Distribution

US Distribution in states of: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, ID, IN, KS, KY, MA, MI, MS, MT, NC, JH, NY, OK, OR, PA, SC, TN, TX, WA, and WI.

Quantity

10,360 plates