FDA Enforcement Class II Terminated

OptiLITE(TM) Products for Laser Surgery, Catalog Numbers: HTM1020F, HTM1040F, HTM1060F, HTM1010F; Manufactured by CONVERGENT LASTER TECHNOLOGIES The Odyssey 30B is approved for numerous medical procedures. Generally, the primary use of the laser (which includes the fibers) is related to urology, but ultimately it is up to the surgeon except in cases where it is contraindicated. The fibers are usually used in cases involving stones anywhere from the bladder up to the kidneys. Occasionally the laser system is used in soft tissue cases (non-cancerous tumors) in and around the urinary system.

Recall: Z-2866-2017 · Reported August 9, 2017

Enforcement

Recall Number
Z-2866-2017
Event ID
77653
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Xintec Corp
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 9, 2017
Initiation Date
June 23, 2017
Classification Date
August 2, 2017
Termination Date
June 20, 2018
Address
1660 S Loop Rd, N/A, Alameda, CA, 94502-7091, United States

Description

OptiLITE(TM) Products for Laser Surgery, Catalog Numbers: HTM1020F, HTM1040F, HTM1060F, HTM1010F; Manufactured by CONVERGENT LASTER TECHNOLOGIES The Odyssey 30B is approved for numerous medical procedures. Generally, the primary use of the laser (which includes the fibers) is related to urology, but ultimately it is up to the surgeon except in cases where it is contraindicated. The fibers are usually used in cases involving stones anywhere from the bladder up to the kidneys. Occasionally the laser system is used in soft tissue cases (non-cancerous tumors) in and around the urinary system.

Reason

It has been determined that the reprocessing instructions may not provide sufficiently detailed information for the cleaning, disinfection and sterilization of the products listed below. The preliminary investigation indicates that validation data related to the reprocessing of these devices may not meet newly enacted standards.

Code Info

all codes.

Distribution

Worldwide distribution- US (Nationwide) and countries of: Canada and Sri Lanka.

Quantity

4343 units