FDA Enforcement
Class II
Terminated
DigitalDiagnost Classic 1.x Upgrades (Stitching Patient Support) 712081
Recall: Z-2856-2018
·
Reported August 29, 2018
Enforcement
- Recall Number
- Z-2856-2018
- Event ID
- 80577
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Medical Systems Gmbh, DMC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 29, 2018
- Initiation Date
- June 27, 2018
- Classification Date
- August 17, 2018
- Termination Date
- April 30, 2024
- Address
- Development And Manufacturing Ctr., Rontgenstr. 24, Hamburg, N/A, Germany
Description
DigitalDiagnost Classic 1.x Upgrades (Stitching Patient Support) 712081
Reason
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Code Info
Serial Numbers: 456897/SN10000166 402837/SN0802410 2010345 410774/SN0802694
Distribution
US Nationwide; International to 66 countries