Broselow/Hinkle, PEDIATRIC EMERGENCY SYSTEM, Broselow Complete Set of Colored Pouches, Product Code 7730MOD
Enforcement
- Recall Number
- Z-2847-2016
- Event ID
- 74647
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Carefusion 2200 Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 28, 2016
- Initiation Date
- June 29, 2016
- Classification Date
- September 20, 2016
- Termination Date
- March 2, 2018
- Address
- 75 N Fairway Dr, Vernon Hills, IL, 60061-1845, United States
Description
Broselow/Hinkle, PEDIATRIC EMERGENCY SYSTEM, Broselow Complete Set of Colored Pouches, Product Code 7730MOD
The Oral Airway pouch within the PINK/RED Oxygen Delivery Module, 7700RAW, was incorrectly packaged with the 8 cm Berman Oral Airway. The pouch contained within the module is correctly labeled as 5 cm but actually contains an 8 cm Berman Oral Airway component. Therefore, the component within the pouch is incorrect. The PINK/RED Oxygen Delivery Module, 7700RAW, is intended to be used within the PINK/RED Pouch of the BROSELOW pediatric safety system. Three (3) lots (lot# 914873, lot# 914876 and lot #914878) of the PINK/RED Oxygen Delivery Module, 7700RAW, are affected by this voluntary Field Safety Corrective Action.
Lot numbers 914873, 914876, and 914878 of the individual modules (7700RAW), which are contained in Lot number 916714 of the Broselow Complete Set of Colored Pouches, Product Code 7730MOD. Expiration Date MAR 2018.
Illinois
50 units