FDA Enforcement Class II Terminated

3D TOP Ceiling Stand, model no. 3070039 3074486 3070013 3070021 Product Usage: The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.

Recall: Z-2837-2018 · Reported August 29, 2018

Enforcement

Recall Number
Z-2837-2018
Event ID
80735
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 29, 2018
Initiation Date
June 29, 2018
Classification Date
August 17, 2018
Termination Date
October 4, 2019
Address
40 Liberty Blvd, Malvern, PA, 19355-1418, United States

Description

3D TOP Ceiling Stand, model no. 3070039 3074486 3070013 3070021 Product Usage: The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.

Reason

In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.

Code Info

3070039: S/N 2271 to 2793 3074486: S/N 1812 to 3036 3070013: S/N 3967 to 5042 3070021: S/N 3852 to 4719

Distribution

US Nationwide Distribution to accounts in IL, FL, and NY.

Quantity

699