FDA Enforcement
Class II
Terminated
Salivary Estradiol ELISA, REF SLV-4188 in vitro diagnostic quantitative measurement of active free Estradiol, an estrogenic steroid, in saliva.
Recall: Z-2836-2018
·
Reported August 22, 2018
Enforcement
- Recall Number
- Z-2836-2018
- Event ID
- 80575
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- DRG Instruments GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 22, 2018
- Initiation Date
- July 14, 2016
- Classification Date
- August 16, 2018
- Termination Date
- August 4, 2020
- Address
- Frauenbergstr. 18, Marburg, N/A, Germany
Description
Salivary Estradiol ELISA, REF SLV-4188 in vitro diagnostic quantitative measurement of active free Estradiol, an estrogenic steroid, in saliva.
Reason
Lower recovery of salivary estradiol levels due to poor differentiation between kit Standard 0 and Standard 1.
Code Info
Lot # 68K056 Exp. Date 2017-05
Distribution
NJ; Germany, Belgium, Spain & Germany
Quantity
169 kits