The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
Recall: Z-2835-2018
·
Reported August 22, 2018
Enforcement
- Recall Number
- Z-2835-2018
- Event ID
- 80691
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Intuitive Surgical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 22, 2018
- Initiation Date
- May 17, 2018
- Classification Date
- August 16, 2018
- Address
- 1266 Kifer Rd Bldg 101, Sunnyvale, CA, 94086-5304, United States
Description
The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
Reason
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
Code Info
UDI Number: 00886874114193. Serial Number: SK1703.
Distribution
U.S.
Quantity
1