FDA Enforcement Class II Terminated

NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.

Recall: Z-2828-2020 · Reported August 26, 2020

Enforcement

Recall Number
Z-2828-2020
Event ID
86085
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
NOBEL BIOCARE SERVICES AG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 26, 2020
Initiation Date
July 20, 2020
Classification Date
August 20, 2020
Termination Date
May 12, 2021
Address
Balz-zimmermannstrasse 7, N/A, Kloten, N/A, N/A, Switzerland

Description

NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.

Reason

Dental implant cap label may incorrectly indicate implant measurements 5.0x8mm. Actual implant measurements are 4.3x10mm. If not noticed prior to implant placement, this could lead to: 1) A second surgical procedure; 2) Implant replacement; 3) If placed in posterior lower jaw, the implant may damage the inferior alveolar nerve canal or perforate a small artery requiring immediate implant removal.

Code Info

Lot: 13084072, UDI: (01)07332747073584(10)13084072(17)240525

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of FL and NY and the countries of Albania and Hong Kong.

Quantity

188