FDA Enforcement Class II Terminated

Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System Product Usage: Stretcher for pre-hospital patient transport.

Recall: Z-2826-2015 · Reported October 7, 2015

Enforcement

Recall Number
Z-2826-2015
Event ID
71983
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ferno-Washington Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
October 7, 2015
Initiation Date
August 3, 2015
Classification Date
September 30, 2015
Termination Date
January 20, 2016
Address
70 Weil Way, N/A, Wilmington, OH, 45177-9371, United States

Description

Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System Product Usage: Stretcher for pre-hospital patient transport.

Reason

Hole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.

Code Info

Serial #s: 15F001703, 15F001704, 15F001706, 15F001707, 15F001708, 15F001709, 15F001710, 15F001711, 15F001712, 15F001713, 15F001715, 15F001717, 15F001719, 15F001721, 15F001722, 15F001723, 15F001724, 15F001725, 15F001726, 15F001727, 15F001728, 15F001729, 15F001730, 15F001731, 15F001732, 15F001733, 15F001734, 15F001735, 15F001736, 15F001737, 15F001738, 15F001739, 15F001740, 15F001741, 15F001742, 15F001744, 15F001745, 15F001746, 15F001747, 15F001748, 15F001749, 15F001750, 15F001751, 15F001752, 15F001753, 15F001754, 15F001755, 15F001756, 15F001757, 15F001758, 15F001759, 15F001760, 15F001761, 15F001762, 15F001763, 15F001764, 15F001765, 15F001766, 15F001767, 15F001768, 15F001769, , 5F001770, 15F001771, 15F001772, 15F001773, 15F001774, 15F001775, 15F001776, 15F001777, 15F001778, 15F001779, 15F001780, 15F001781, 15F001782, 15F001783, 15F001784, 15F001785, 15F001786, 15F001787, 15F001788, 15F001789, 15F001790, 15F001791, 15F001793, 15F001794, 15F001795, 15F001796, 15F001797, 15F001798, 15F001804, 15F001805, 15F001806, 15F001807, 15F001808, 15F001809, 15F001810, 15F001811, 15F001812, 15F001813, 15F001814, 15F001815, 15F001816 & 15F001850.

Distribution

Worldwide Distribution - US Nationwide in the states of: CO, IA, IL, IN, KY, MD, MN, MO, NC, NJ, NY, OH, OK, OR, TX, VA, WA & WY and the countries of CANADA, AUSTRALIA, UNITED KINGDOM & NEW ZEALAND.

Quantity

72 units