FDA Enforcement
Class II
Terminated
ASP-1020 LipoTower with Aspiration Pump. Medical use.
Recall: Z-2821-2015
·
Reported October 7, 2015
Enforcement
- Recall Number
- Z-2821-2015
- Event ID
- 72106
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- MicroAire Surgical Instruments, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- October 7, 2015
- Initiation Date
- June 12, 2015
- Classification Date
- September 30, 2015
- Termination Date
- May 23, 2016
- Address
- 3590 Grand Forks Blvd, N/A, Charlottesville, VA, 22911-9006, United States
Description
ASP-1020 LipoTower with Aspiration Pump. Medical use.
Reason
There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails.
Code Info
Serial numbers 14031102, 15021105, and 14101103.
Distribution
Nationwide Distribution.
Quantity
3 LipoTower with Aspiration Pump