FDA Enforcement Class II Terminated

DePuy Synthes Trauma LCP One-Third Tubular Plate with Collar 8 Holes / 93mm - Product Usage: intended for fixation of fractures, osteotomies and non-unions of clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia and fibula, particularly in osteopenic bone. Part Number 241.381

Recall: Z-2816-2020 · Reported August 26, 2020

Enforcement

Recall Number
Z-2816-2020
Event ID
86104
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Synthes (USA) Products LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 26, 2020
Initiation Date
July 21, 2020
Classification Date
August 17, 2020
Termination Date
February 18, 2022
Address
1301 Goshen Pkwy, N/A, West Chester, PA, 19380-5986, United States

Description

DePuy Synthes Trauma LCP One-Third Tubular Plate with Collar 8 Holes / 93mm - Product Usage: intended for fixation of fractures, osteotomies and non-unions of clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia and fibula, particularly in osteopenic bone. Part Number 241.381

Reason

May have one or more dimensional features out of tolerance includes: thread pitch diameter depth, thread minor diameter depth, plate thickness and end length out of tolerance that can potentially result in a reduced locking strength of an LCP screw

Code Info

Lot Number: 29P6036 UDI: 10886982166913

Distribution

US Nationwide distribution.

Quantity

191 units