Siemens RAPIDPoint 500 Measurement Cartridge Lactate (400) tests Part Number: 10491448 Product Usage: The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCHb, FMetHb, FHHb, nBili
Enforcement
- Recall Number
- Z-2814-2015
- Event ID
- 72044
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 30, 2015
- Initiation Date
- August 24, 2015
- Classification Date
- September 23, 2015
- Termination Date
- May 23, 2016
- Address
- 2 Edgewater Dr, N/A, Norwood, MA, 02062-4637, United States
Description
Siemens RAPIDPoint 500 Measurement Cartridge Lactate (400) tests Part Number: 10491448 Product Usage: The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCHb, FMetHb, FHHb, nBili
RAPIDPoint 500 measurement cartridge serial numbers 2517102517 through 2519000012, negative sodium bias on both QC and patient results
Serial Numbers: 2517102517 through 2519000012. Cartridges within this serial number range that have a green dot on the box and/or the cartridge are not affected
Worldwide Distribution - US Nationwide and the countries of Canada, Albania, Austria, Australia, Bosnia Herzeg., Bangladesh, Belgium, Brazil, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Denmark, Estonia, Spain, Finland, France, Great Britain, Georgia, Greece, Croatia, Hungary, Ireland, India, Italy, Lithuania, Luxembourg, Latvia, Macedonia, Malaysia, Netherlands, Norway, New Zealand, Poland, Portugal, Paraguay, Romania, Sweden, Singapore, Slovenia, Slovakia, Turkey, Taiwan, South Africa, Unit.Arab Emirates
1213 cartridges