FDA Enforcement Class II Terminated

Phillips 3.7V Rechargeable Li-Polymer Battery, Product #: 989803191341 - Product Usage: This battery is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.

Recall: Z-2810-2020 · Reported August 26, 2020

Enforcement

Recall Number
Z-2810-2020
Event ID
86025
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips North America, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 26, 2020
Initiation Date
June 18, 2020
Classification Date
August 14, 2020
Termination Date
January 10, 2023
Address
3000 Minuteman Rd, N/A, Andover, MA, 01810-1032, United States

Description

Phillips 3.7V Rechargeable Li-Polymer Battery, Product #: 989803191341 - Product Usage: This battery is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.

Reason

A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.

Code Info

Product #: 989803191341 Manufacturing date codes 2018-12-DD to 2020-02-DD

Distribution

World wide distribution. Firm reports distributing more product then quantity manufactured. Clarification requested. And received. Firm made error in counting number of batteries distributed.

Quantity

24,545 Units