The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica ASP6025 Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.
Enforcement
- Recall Number
- Z-2804-2015
- Event ID
- 71928
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Leica Microsystems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 30, 2015
- Initiation Date
- July 27, 2015
- Classification Date
- September 22, 2015
- Termination Date
- January 5, 2017
- Address
- 1700 Leider Ln, N/A, Buffalo Grove, IL, 60089-6622, United States
Description
The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica ASP6025 Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.
The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.
Product Code: Leica ASP6025 Serial Numbers: 163, 169, 200-729, 731, 733, 735, 737, 739, 741, 743, 745, 747, 749, 751, 753, 755, 757, 759, 761, 763, 765, 767, 769, 771, 773, 775, 777, 779, 781, 783, 785, 787, 789, 791, 793, 795, 797, 799, 801, 803, 805, 807, 809, 811.
Worldwide Distribution - US including AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH,PA, SC, TN, TX, VA, WA, WI, WV and District of Columbia. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Switzerland, Chile, China, Colombia, Czech Republic , Germany, Spain, France, England, Georgia, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, Korea, Kuwait, Lithuania, Myanmar, Mexico, Malaysia, Netherland, Norway, Peru, Philippines, Poland, Qatar, Reunion, Romania, Serbia, Saudi Arabia, Singapore, Slovenia, Thailand, Turkey, Taiwan, and South Africa.
U.S. Quantity Distributed: 163 systems; Foreign Quantity Diistributed: 410 systems.