uMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Enforcement
- Recall Number
- Z-2800-2020
- Event ID
- 85948
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 19, 2020
- Initiation Date
- July 16, 2020
- Classification Date
- August 13, 2020
- Termination Date
- July 7, 2022
- Address
- No. 2258 Chengbei Rd, Jiading Ind., N/A, Shanghai, N/A, N/A, China
Description
uMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
1) Service function problem may cause false marking of a bad channel resulting in ring artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions (N/A for Serial No. 200036)
Serial No.s 200017, 200023, 200024 and 200036.
US Nationwide distribution including in the state of Texas.
4 units