FDA Enforcement Class II Terminated

Bioseal Trachea Extender 1/pl 50pk/Cs, REF TRAX01/50, LOT XXXX, STERILE EO, Single Use Only, Date of Manufacture XXXX, Expiration Date XXXX, LF, UDI: (01) 00630094916019 - Product Usage: Airway connectors intended to connect a breathing gas source to a tracheal tube, tracheostomy tube, or mask.

Recall: Z-2795-2020 · Reported August 19, 2020

Enforcement

Recall Number
Z-2795-2020
Event ID
85288
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Bioseal Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 19, 2020
Initiation Date
August 15, 2019
Classification Date
August 13, 2020
Termination Date
March 4, 2021
Address
167 W Orangethorpe Ave, N/A, Placentia, CA, 92870-6922, United States

Description

Bioseal Trachea Extender 1/pl 50pk/Cs, REF TRAX01/50, LOT XXXX, STERILE EO, Single Use Only, Date of Manufacture XXXX, Expiration Date XXXX, LF, UDI: (01) 00630094916019 - Product Usage: Airway connectors intended to connect a breathing gas source to a tracheal tube, tracheostomy tube, or mask.

Reason

The reported stress crack in the port of the swivel elbow where the flip cap is inserted could result in leakage from the breathing system resulting in the prescribed ventilation not being delivered to the patient. These cracks could possibly leak if they reach a significant size and pose a serious heath risk to patient.

Code Info

Bioseal Part # TRAX01/50 Bioseal Lot #'s 400453, 400456, 400461, 400472, 400476

Distribution

US Nationwide distribution including in the state of CA.

Quantity

900 units