FDA Enforcement Class II Terminated

Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure ventilation support to adults and pediatrics. The device is used in the in the intensive care ward, in the recovery room, and during transfer of ventilated patients within the hospital.

Recall: Z-2795-2018 · Reported August 22, 2018

Enforcement

Recall Number
Z-2795-2018
Event ID
80467
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hamilton Medical AG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 22, 2018
Initiation Date
March 9, 2018
Classification Date
August 13, 2018
Termination Date
March 18, 2019
Address
Via Crusch 8, Bonaduz, N/A, Switzerland

Description

Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure ventilation support to adults and pediatrics. The device is used in the in the intensive care ward, in the recovery room, and during transfer of ventilated patients within the hospital.

Reason

An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.

Code Info

Serial Number: 12244, 12246, 12338, 12343, 12344, 12345, 12351, 12352, 12355, 12370, 12380, 12382, 12383, 12412, 12426, 12432

Distribution

US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.

Quantity

16 devices