Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure ventilation support to adults and pediatrics. The device is used in the in the intensive care ward, in the recovery room, and during transfer of ventilated patients within the hospital.
Enforcement
- Recall Number
- Z-2795-2018
- Event ID
- 80467
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Hamilton Medical AG
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 22, 2018
- Initiation Date
- March 9, 2018
- Classification Date
- August 13, 2018
- Termination Date
- March 18, 2019
- Address
- Via Crusch 8, Bonaduz, N/A, Switzerland
Description
Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure ventilation support to adults and pediatrics. The device is used in the in the intensive care ward, in the recovery room, and during transfer of ventilated patients within the hospital.
An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.
Serial Number: 12244, 12246, 12338, 12343, 12344, 12345, 12351, 12352, 12355, 12370, 12380, 12382, 12383, 12412, 12426, 12432
US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.
16 devices