AccuLIF PL 10-16mm x 11mm x 27mm x 8 Cage, Rx only, Sterile R, Legal The AccuLIF PL Cages are indicated for intervertebral body fusion with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the AccuLIF PL Cages can be used as in adjunct to fusion in patients diagnosed with degenerative scoliosis. The AccuLIF Pl Cages are always to be used with supplemental internal spinal fixation. Additionally, the AccuLIF PL Cages are to be used with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion
Enforcement
- Recall Number
- Z-2788-2016
- Event ID
- 74931
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Howmedica Osteonics Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 21, 2016
- Initiation Date
- July 26, 2016
- Classification Date
- September 12, 2016
- Termination Date
- March 12, 2018
- Address
- 2 Pearl Ct, N/A, Allendale, NJ, 07401-1611, United States
Description
AccuLIF PL 10-16mm x 11mm x 27mm x 8 Cage, Rx only, Sterile R, Legal The AccuLIF PL Cages are indicated for intervertebral body fusion with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the AccuLIF PL Cages can be used as in adjunct to fusion in patients diagnosed with degenerative scoliosis. The AccuLIF Pl Cages are always to be used with supplemental internal spinal fixation. Additionally, the AccuLIF PL Cages are to be used with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion
Since Stryker acquired the AccuLIF product in March of 2014, there have been Product Inquires (PIs) related to the PL implant confirming to have a reduction of height > 1 mm. The current occurrence rate for reported post-operative reduction of cage height (> 1mm) of AccuLIF PL Expandable Cages is approximately 0.47% and has demonstrated a slight upward trend since September 2015.
Catalog #400010 Lot #'s 01021403, 01121508, 01121509, 01271515, 01271516, 02021512, 02021513, 02021514, 02021515, 02021602, 03221310, 04141508, 04141509, 04141510, 04141511, 04141512, 04251312, 06101405, 06101406, 06101407, 06251311, 07011406, 07201510, 07201511, 07201512, 08141303, 08141313, 09251504, 09281512, 10081301, 10081302, 10201502, 10201503, 10301504, 11121514, 11121518, 11121521, 12151511 and 12151517
NationwideDistribution