FDA Enforcement
Class II
Terminated
SORIN GROUP, SMART PERFUSION PACK, INSPIRE 6FS, REF 044008204, STERILE EO, Rx Only, 1 EA
Recall: Z-2786-2017
·
Reported August 9, 2017
Enforcement
- Recall Number
- Z-2786-2017
- Event ID
- 77810
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Sorin Group USA, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- August 9, 2017
- Initiation Date
- June 29, 2017
- Classification Date
- July 29, 2017
- Termination Date
- April 12, 2021
- Address
- 14401 W 65th Way, N/A, Arvada, CO, 80004-3503, United States
Description
SORIN GROUP, SMART PERFUSION PACK, INSPIRE 6FS, REF 044008204, STERILE EO, Rx Only, 1 EA
Reason
Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
Code Info
Lot Numbers: 1626400126, 1628100032, 1631900034, 1632800034, 1703300043, 1707400063, 1708000042, 1710000040, 1716500018
Distribution
Nationwide
Quantity
52 units