FDA Enforcement
Class II
Ongoing
Centurion procedure packs, containing Caina syringes, labeled as: 1) NICU CIRCUMCISION TRAY W/OUT CLAMP, Pack Number CIT6920; 2) RETINAL INJECTION KIT, Pack Number MNS12465
Recall: Z-2774-2024
·
Reported September 4, 2024
Enforcement
- Recall Number
- Z-2774-2024
- Event ID
- 94586
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- MEDLINE INDUSTRIES, LP - Northfield
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- September 4, 2024
- Initiation Date
- April 8, 2024
- Classification Date
- August 28, 2024
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Centurion procedure packs, containing Caina syringes, labeled as: 1) NICU CIRCUMCISION TRAY W/OUT CLAMP, Pack Number CIT6920; 2) RETINAL INJECTION KIT, Pack Number MNS12465
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Code Info
CIT6920, Lot Number 2021092290; MNS12465, Lot Number 2022040190
Distribution
Worldwide distribution.