Voyant Open Fusion Device Models: EB040 and EB040+ Product Usage: The Applied Voyant Open Fusion Device is designed for use with the Voyant Electrosurgical Generator. The device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 5mm (EB040+) or 7mm (EB040) in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The system delivers the energy required to fuse the tissue between the jaws of the device by monitoring voltage and current throughout the seal cycle. The device also features a useractuated blade for the transection of sealed tissue.
Enforcement
- Recall Number
- Z-2769-2015
- Event ID
- 71956
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Applied Medical Resources Corp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 23, 2015
- Initiation Date
- July 31, 2015
- Classification Date
- September 15, 2015
- Termination Date
- November 20, 2015
- Address
- 22872 Avenida Empresa, Rancho Santa Margarita, CA, 92688-2650, United States
Description
Voyant Open Fusion Device Models: EB040 and EB040+ Product Usage: The Applied Voyant Open Fusion Device is designed for use with the Voyant Electrosurgical Generator. The device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 5mm (EB040+) or 7mm (EB040) in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The system delivers the energy required to fuse the tissue between the jaws of the device by monitoring voltage and current throughout the seal cycle. The device also features a useractuated blade for the transection of sealed tissue.
Potential to improperly characterize tissue within the jaws, which may result in nonoptimal power delivery.
Model Number: EB040 Lot Numbers: 1243964, 1243975, 1244125 Model Number: EB040+ Lot Numbers: 1245512, 1245513, 1246241, 1246508, 1247079, 1247566, 1248102, 1249264
Worldwide Distribution - U.S Nationwide in the states of:FL, GA, NV. and countries of: France, Netherlands, Spain, and United Kingdom.
101 units (U.S) and 38 units (International)