FDA Enforcement Class II Terminated

MedTest DX Control Reconstitution Fluid Kit configuration: 20 x 5 ml. Bottle: 5 ml amber glass Cap: 18 mm white seal plug seal rubber HC-DIL (Human Control Diluent) is reagent grade water filled into 5ml vials used to reconstitute Human controls HNC200 and HEC200. in vitro diagnostic.

Recall: Z-2768-2015 · Reported September 23, 2015

Enforcement

Recall Number
Z-2768-2015
Event ID
71993
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
MedtestDx, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 23, 2015
Initiation Date
May 15, 2015
Classification Date
September 15, 2015
Termination Date
May 23, 2016
Address
5449 Research Dr, Canton, MI, 48188-2261, United States

Description

MedTest DX Control Reconstitution Fluid Kit configuration: 20 x 5 ml. Bottle: 5 ml amber glass Cap: 18 mm white seal plug seal rubber HC-DIL (Human Control Diluent) is reagent grade water filled into 5ml vials used to reconstitute Human controls HNC200 and HEC200. in vitro diagnostic.

Reason

Vial to vial variation in the fill volume that could affect control recovery issues and cause a control result to fall outside the suggested recovery ranges. An under-fill of the vials was found ranging as low as 4.63 ml. When used to reconstitute HNC/HEC, control values were shown to be out of range running on the high end.

Code Info

Catalog number: HC-DIL Lot: 072414-206 Expiration date: 2016-07

Distribution

US Nationwide Distribution.

Quantity

79 kits