NICO Myriad Console Model number NN-7000 The Myriad System is a surgical system specifically designed for morcellation of soft tissue. It primarily consists of a reusable console with foot pedal and a disposable handpiece. For the morcellation and removal of tissue during pelviscopic, laparoscopic, percutaneous and open surgical procedures whenever access to the surgical site is limited.
Enforcement
- Recall Number
- Z-2767-2015
- Event ID
- 72015
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Nico Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- FAX
- Report Date
- September 23, 2015
- Initiation Date
- July 29, 2015
- Classification Date
- September 15, 2015
- Termination Date
- September 22, 2015
- Address
- 250 E 96th St, ste 125, Indianapolis, IN, 46240-3781, United States
Description
NICO Myriad Console Model number NN-7000 The Myriad System is a surgical system specifically designed for morcellation of soft tissue. It primarily consists of a reusable console with foot pedal and a disposable handpiece. For the morcellation and removal of tissue during pelviscopic, laparoscopic, percutaneous and open surgical procedures whenever access to the surgical site is limited.
Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.
Model number NN-7000 SN: 00231, 00232, 00242, 00245, 00248, 00250
US Distribution to states of: MD, IN, CA, and NC.
6 units