Medline procedural kits labeled as: 1) ALL DOC EYE PACK, Pack Number DYNJ51654A ; 2) ASC CATARACT PACK , Pack Number DYNJ0855117O ; 3) BAI , Pack Number DYNJ53847D ; 4) BASIC EYE PACK, Pack Number DYNJ83699A ; 5) BASIC EYE PACK, Pack Number DYNJ81805; 6) BASIC EYE PACK, Pack Number DYNJ83699A ; 7) BYERS CATARACT, Pack Number DYNJ67344; 8) BYERS CATARACT, Pack Number DYNJ67344H ; 9) CATARACT, Pack Number DYNJ44544B ; 10) CATARACT, Pack Number DYNJ54713C ; 11) CATARACT CDS , Pack Number CDS983804F ; 12) CATARACT CDS, Pack Number CDS983804F ; 13) CATARACT DR MALONE PACK , Pack Number DYNJ51590A ; 14) CATARACT DR MULROONEY PACK, Pack Number DYNJ51595A ; 15) CATARACT DR STEVENS PACK, Pack Number DYNJ51593C ; 16) CATARACT L J, Pack Number DYNJ40056A ; 17) CATARACT PACK, Pack Number CMPJ03235C ; 18) CATARACT PACK, Pack Number DYNJ80316A ; 19) CATARACT PACK, Pack Number VAL011CTASF; 20) CATARACT PACK , Pack Number CMPJ03235C ; 21) CATARACT PACK , Pack Number DYNJ31020K ; 22) CATARACT PACK , Pack Number DYNJ31020L ; 23) CATARACT PACK , Pack Number DYNJ54790D ; 24) CATARACT PACK , Pack Number DYNJ64962; 25) CATARACT PACK , Pack Number DYNJ80316; 26) CATARACT PACK , Pack Number DYNJ80316A ; 27) CATARACT PACK , Pack Number DYNJ80487A ; 28) CATARACT PACK , Pack Number DYNJ80487B ; 29) CATARACT PACK , Pack Number DYNJ80487C ; 30) CATARACT PACK , Pack Number DYNJ80487F ; 31) CATARACT PACK , Pack Number DYNJ84265A ; 32) CATARACT PACK , Pack Number DYNJ85332; 33) CATARACT PACK , Pack Number LYN042CPFEG; 34) CATARACT PACK , Pack Number VAL011CTASF; 35) CATARACT PACK , Pack Number DYNJ43048K ; 36) CATARACT PACK , Pack Number DYNJ52649D ; 37) CATARACT PACK , Pack Number DYNJ52649F ; 38) CATARACT PACK , Pack Number DYNJ55682D ; 39) CATARACT PACK , Pack Number DYNJ63284A ; 40) CATARACT PACK , Pack Number DYNJ63284B ; 41) CATARACT PACK , Pack Number DYNJ66410B ; 42) CATARACT PACK , Pack Number DYNJ85042; 43) CATARACT PACK , Pack Number DYNJ68836B ; 44) CATARACT PACK - GB, Pack Number DYNJ54790F ; 45) CATARACT PACK LUTHER HOSPITAL , Pack Number DYNJ61699D ; 46) CATARACT PACK WITH NO TUB , Pack Number DYNJ80232; 47) CATARACT PACK WITH NO TUB , Pack Number DYNJ80232H ; 48) CATARACT PACK-LF, Pack Number DYNJ42830D ; 49) CATARACT PACK-LF, Pack Number DYNJ0200004B ; 50) CATARACT PACK-LF, Pack Number DYNJ36100C ; 51) CATARACT TRAY , Pack Number DYNJ44356I ; 52) CATARACT TRAY , Pack Number DYNJ44356J ; 53) CATARACT TRAY , Pack Number DYNJ62410A ; 54) COVENANT RETINAL TRAY , Pack Number EYT102K; 55) CUSTOM EYE TRAY #1, Pack Number DYNJ66003D ; 56) CUSTOM EYE TRAY #1, Pack Number DYNJ66003C ; 57) CUSTOM EYE TRAY #1, Pack Number DYNJ66003D ; 58) CUSTOM EYE TRAY #2 , Pack Number DYNJ66004B ; 59) CUSTOM EYE TRAY #2, Pack Number DYNJ66004A ; 60) CUSTOM EYE TRAY #2, Pack Number DYNJ66004B ; 61) DDD VITRECTOMY PACK-LF, Pack Number DYNJ19973S ; 62) DR FLOOD EYE PACK , Pack Number DYNJ57258G ; 63) DR. OSTER GROSSER EYE PACK, Pack Number DYNJ44671C ; 64) DR. OSTER GROSSER EYE PACK, Pack Number DYNJ44671D ; 65) EYE , Pack Number DYNJ0559863K ; 66) EYE BCS PACK, Pack Number DYNJ37682J ; 67) EYE BCS PACK, Pack Number DYNJ37682K ; 68) EYE CATARACT (BASIC EYE), Pack Number DYNJ63272B ; 69) EYE INJECTION KIT , Pack Number MNS13080 ; 70) EYE MUSCLE PACK , Pack Number DYNJ66022A ; 71) EYE MUSCLE TRAY FFX , Pack Number DYNJ63638A ; 72) EYE PACK, Pack Number DYNJ40135C ; 73) EYE PACK, Pack Number DYNJ55484A ; 74) EYE PACK, Pack Number DYNJ60960A ; 75) EYE PACK, Pack Number DYNJ67973B ; 76) EYE PACK, Pack Number DYNJ83799C ; 77) EYE PACK, Pack Number VAL013EPWEC; 78) EYE PACK, Pack Number VAL038BSCLC; 79) EYE PACK, Pack Number DYNJ56293D ; 80) EYE PACK, Pack Number DYNJ64051C ; 81) EYE PACK, Pack Number CMPJ06528F ; 82) EYE PACK, Pack Number DYNJ38400I ; 83) EYE PACK, Pack Number DYNJ38400J ; 84) EYE PACK, Pack Number DYNJ64284C ; 85) EYE PACK, Pack Number DYNJ64284D ; 86) EYE PACK, Pack Numbe
Enforcement
- Recall Number
- Z-2764-2024
- Event ID
- 94581
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- MEDLINE INDUSTRIES, LP - Northfield
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- September 4, 2024
- Initiation Date
- April 8, 2024
- Classification Date
- August 28, 2024
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline procedural kits labeled as: 1) ALL DOC EYE PACK, Pack Number DYNJ51654A ; 2) ASC CATARACT PACK , Pack Number DYNJ0855117O ; 3) BAI , Pack Number DYNJ53847D ; 4) BASIC EYE PACK, Pack Number DYNJ83699A ; 5) BASIC EYE PACK, Pack Number DYNJ81805; 6) BASIC EYE PACK, Pack Number DYNJ83699A ; 7) BYERS CATARACT, Pack Number DYNJ67344; 8) BYERS CATARACT, Pack Number DYNJ67344H ; 9) CATARACT, Pack Number DYNJ44544B ; 10) CATARACT, Pack Number DYNJ54713C ; 11) CATARACT CDS , Pack Number CDS983804F ; 12) CATARACT CDS, Pack Number CDS983804F ; 13) CATARACT DR MALONE PACK , Pack Number DYNJ51590A ; 14) CATARACT DR MULROONEY PACK, Pack Number DYNJ51595A ; 15) CATARACT DR STEVENS PACK, Pack Number DYNJ51593C ; 16) CATARACT L J, Pack Number DYNJ40056A ; 17) CATARACT PACK, Pack Number CMPJ03235C ; 18) CATARACT PACK, Pack Number DYNJ80316A ; 19) CATARACT PACK, Pack Number VAL011CTASF; 20) CATARACT PACK , Pack Number CMPJ03235C ; 21) CATARACT PACK , Pack Number DYNJ31020K ; 22) CATARACT PACK , Pack Number DYNJ31020L ; 23) CATARACT PACK , Pack Number DYNJ54790D ; 24) CATARACT PACK , Pack Number DYNJ64962; 25) CATARACT PACK , Pack Number DYNJ80316; 26) CATARACT PACK , Pack Number DYNJ80316A ; 27) CATARACT PACK , Pack Number DYNJ80487A ; 28) CATARACT PACK , Pack Number DYNJ80487B ; 29) CATARACT PACK , Pack Number DYNJ80487C ; 30) CATARACT PACK , Pack Number DYNJ80487F ; 31) CATARACT PACK , Pack Number DYNJ84265A ; 32) CATARACT PACK , Pack Number DYNJ85332; 33) CATARACT PACK , Pack Number LYN042CPFEG; 34) CATARACT PACK , Pack Number VAL011CTASF; 35) CATARACT PACK , Pack Number DYNJ43048K ; 36) CATARACT PACK , Pack Number DYNJ52649D ; 37) CATARACT PACK , Pack Number DYNJ52649F ; 38) CATARACT PACK , Pack Number DYNJ55682D ; 39) CATARACT PACK , Pack Number DYNJ63284A ; 40) CATARACT PACK , Pack Number DYNJ63284B ; 41) CATARACT PACK , Pack Number DYNJ66410B ; 42) CATARACT PACK , Pack Number DYNJ85042; 43) CATARACT PACK , Pack Number DYNJ68836B ; 44) CATARACT PACK - GB, Pack Number DYNJ54790F ; 45) CATARACT PACK LUTHER HOSPITAL , Pack Number DYNJ61699D ; 46) CATARACT PACK WITH NO TUB , Pack Number DYNJ80232; 47) CATARACT PACK WITH NO TUB , Pack Number DYNJ80232H ; 48) CATARACT PACK-LF, Pack Number DYNJ42830D ; 49) CATARACT PACK-LF, Pack Number DYNJ0200004B ; 50) CATARACT PACK-LF, Pack Number DYNJ36100C ; 51) CATARACT TRAY , Pack Number DYNJ44356I ; 52) CATARACT TRAY , Pack Number DYNJ44356J ; 53) CATARACT TRAY , Pack Number DYNJ62410A ; 54) COVENANT RETINAL TRAY , Pack Number EYT102K; 55) CUSTOM EYE TRAY #1, Pack Number DYNJ66003D ; 56) CUSTOM EYE TRAY #1, Pack Number DYNJ66003C ; 57) CUSTOM EYE TRAY #1, Pack Number DYNJ66003D ; 58) CUSTOM EYE TRAY #2 , Pack Number DYNJ66004B ; 59) CUSTOM EYE TRAY #2, Pack Number DYNJ66004A ; 60) CUSTOM EYE TRAY #2, Pack Number DYNJ66004B ; 61) DDD VITRECTOMY PACK-LF, Pack Number DYNJ19973S ; 62) DR FLOOD EYE PACK , Pack Number DYNJ57258G ; 63) DR. OSTER GROSSER EYE PACK, Pack Number DYNJ44671C ; 64) DR. OSTER GROSSER EYE PACK, Pack Number DYNJ44671D ; 65) EYE , Pack Number DYNJ0559863K ; 66) EYE BCS PACK, Pack Number DYNJ37682J ; 67) EYE BCS PACK, Pack Number DYNJ37682K ; 68) EYE CATARACT (BASIC EYE), Pack Number DYNJ63272B ; 69) EYE INJECTION KIT , Pack Number MNS13080 ; 70) EYE MUSCLE PACK , Pack Number DYNJ66022A ; 71) EYE MUSCLE TRAY FFX , Pack Number DYNJ63638A ; 72) EYE PACK, Pack Number DYNJ40135C ; 73) EYE PACK, Pack Number DYNJ55484A ; 74) EYE PACK, Pack Number DYNJ60960A ; 75) EYE PACK, Pack Number DYNJ67973B ; 76) EYE PACK, Pack Number DYNJ83799C ; 77) EYE PACK, Pack Number VAL013EPWEC; 78) EYE PACK, Pack Number VAL038BSCLC; 79) EYE PACK, Pack Number DYNJ56293D ; 80) EYE PACK, Pack Number DYNJ64051C ; 81) EYE PACK, Pack Number CMPJ06528F ; 82) EYE PACK, Pack Number DYNJ38400I ; 83) EYE PACK, Pack Number DYNJ38400J ; 84) EYE PACK, Pack Number DYNJ64284C ; 85) EYE PACK, Pack Number DYNJ64284D ; 86) EYE PACK, Pack Numbe
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
DYNJ51654A , Lot Number:23HBZ446 ; DYNJ0855117O , Lot Number:23GME420 ; DYNJ0855117O , Lot Number:23IMA529 ; DYNJ0855117O , Lot Number:24AMG373 ; DYNJ53847D , Lot Number:23AMJ016 ; DYNJ53847D , Lot Number:23EMJ451 ; DYNJ53847D , Lot Number:23FMF604 ; DYNJ53847D , Lot Number:23JMI394 ; DYNJ53847D , Lot Number:23KMB963 ; DYNJ53847D , Lot Number:24AMA722 ; DYNJ83699A , Lot Number:24CMG922 ; DYNJ81805, Lot Number:23FMD167 ; DYNJ81805, Lot Number:23GMA680 ; DYNJ81805, Lot Number:23JMD040 ; DYNJ81805, Lot Number:24AMA541 ; DYNJ83699A , Lot Number:23KMA807 ; DYNJ83699A , Lot Number:24AMB925 ; DYNJ67344, Lot Number:23IMH376 ; DYNJ67344, Lot Number:23JMG829 ; DYNJ67344H , Lot Number:23IMH376 ; DYNJ67344H , Lot Number:23JMG829 ; DYNJ44544B , Lot Number:24BBN345 ; DYNJ54713C , Lot Number:23GBR621 ; CDS983804F , Lot Number:24CBG535 ; CDS983804F , Lot Number:23BBN605 ; CDS983804F , Lot Number:23BBU281 ; CDS983804F , Lot Number:23EBC831 ; CDS983804F , Lot Number:23EBW202 ; CDS983804F , Lot Number:23FBH069 ; CDS983804F , Lot Number:23HBD970 ; CDS983804F , Lot Number:23IBC407 ; CDS983804F , Lot Number:23JBF548 ; DYNJ51590A , Lot Number:21KBS483 ; DYNJ51590A , Lot Number:21KBS490 ; DYNJ51590A , Lot Number:22GBD368 ; DYNJ51590A , Lot Number:22JBL806 ; DYNJ51590A , Lot Number:22JBU147 ; DYNJ51590A , Lot Number:22JBW972 ; DYNJ51590A , Lot Number:23FBR199 ; DYNJ51590A , Lot Number:23IBJ742 ; DYNJ51590A , Lot Number:23JBK026 ; DYNJ51595A , Lot Number:21IBN715 ; DYNJ51595A , Lot Number:21KBJ227 ; DYNJ51595A , Lot Number:21LBQ210 ; DYNJ51595A , Lot Number:22FBB062 ; DYNJ51595A , Lot Number:22GBN540 ; DYNJ51595A , Lot Number:22IBU646 ; DYNJ51595A , Lot Number:23ABO675 ; DYNJ51595A , Lot Number:23CBL791 ; DYNJ51595A , Lot Number:23EBA140 ; DYNJ51595A , Lot Number:23IBM004 ; DYNJ51593C , Lot Number:21KBI577 ; DYNJ51593C , Lot Number:21KBT571 ; DYNJ51593C , Lot Number:21LBT302 ; DYNJ51593C , Lot Number:22JBA148 ; DYNJ51593C , Lot Number:23BBK907 ; DYNJ40056A , Lot Number:23HMG522 ; DYNJ40056A , Lot Number:23KMB058 ; CMPJ03235C , Lot Number:24CBJ839 ; DYNJ80316A , Lot Number:24CBF142 ; VAL011CTASF, Lot Number:24CBL957 ; CMPJ03235C , Lot Number:23BBB565 ; CMPJ03235C , Lot Number:23GBO681 ; CMPJ03235C , Lot Number:24BBI378 ; DYNJ31020K , Lot Number:21LBL497 ; DYNJ31020K , Lot Number:22OBC488 ; DYNJ31020K , Lot Number:22FBT609 ; DYNJ31020K , Lot Number:22HBA842 ; DYNJ31020K , Lot Number:23BBB378 ; DYNJ31020K , Lot Number:23CBJ094 ; DYNJ31020K , Lot Number:23DBQ602 ; DYNJ31020K , Lot Number:23FBK551 ; DYNJ31020K , Lot Number:23JBT069 ; DYNJ31020K , Lot Number:20XBG244 ; DYNJ31020K , Lot Number:21EBB373 ; DYNJ31020L , Lot Number:23LBO289 ; DYNJ31020L , Lot Number:23LBS944 ; DYNJ54790D , Lot Number:21KBK564 ; DYNJ54790D , Lot Number:21LBP312 ; DYNJ64962, Lot Number:22EBE357 ; DYNJ64962, Lot Number:22GBM208 ; DYNJ64962, Lot Number:23CBM665 ; DYNJ64962, Lot Number:23DBC274 ; DYNJ64962, Lot Number:23EBD021 ; DYNJ64962, Lot Number:23FBV254 ; DYNJ64962, Lot Number:21ABD721 ; DYNJ64962, Lot Number:21ABG947 ; DYNJ80316, Lot Number:22GBK573 ; DYNJ80316A , Lot Number:23EBX492 ; DYNJ80316A , Lot Number:23KBB868 ; DYNJ80316A , Lot Number:23LBE622 ; DYNJ80316A , Lot Number:24ABA118 ; DYNJ80316A , Lot Number:24ABQ972 ; DYNJ80316A , Lot Number:24BBL220 ; DYNJ80487A , Lot Number:22HBI178 ; DYNJ80487B , Lot Number:23BBR916 ; DYNJ80487B , Lot Number:23CBP361 ; DYNJ80487C , Lot Number:23FBO339 ; DYNJ80487C , Lot Number:23GBO642 ; DYNJ80487C , Lot Number:23GBR276 ; DYNJ80487C , Lot Number:23HBP753 ; DYNJ80487F , Lot Number:23KBF204 ; DYNJ80487F , Lot Number:23KBF967 ; DYNJ80487F , Lot Number:24BBJ804 ; DYNJ84265A , Lot Number:23LBP047 ; DYNJ84265A , Lot Number:24ABU345 ; DYNJ85332, Lot Number:23KBM469 ; LYN042CPFEG, Lot Number:21LBM245 ; VAL011CTASF, Lot Number:23CBI687 ; VAL011CTASF, Lot Number:23DBO423 ; VAL011CTASF, Lot Number:23FBD539 ; VAL011CTASF, Lot Number:23IBE673 ; VAL011CTASF, Lot Number:23KBC654 ; VAL011CTASF, Lot Number:23LBG127 ; DYNJ43048K , Lot Number:23JMF537 ; DYNJ52649D , Lot Number:21CMD757 ; DYNJ52649D , Lot Number:21HMD666 ; DYNJ52649F , Lot Number:23FMJ496 ; DYNJ52649F , Lot Number:23GMC382 ; DYNJ52649F , Lot Number:23HME435 ; DYNJ52649F , Lot Number:23KMF720 ; DYNJ52649F , Lot Number:23LMG694 ; DYNJ54790D , Lot Number:23AMB680 ; DYNJ55682D , Lot Number:23JMB655 ; DYNJ63284A , Lot Number:21HMA782 ; DYNJ63284A , Lot Number:22DME834 ; DYNJ63284A , Lot Number:23CMA698 ; DYNJ63284B , Lot Number:23GMH292 ; DYNJ63284B , Lot Number:23KMB077 ; DYNJ63284B , Lot Number:24AMB323 ; DYNJ64962, Lot Number:23KMH920 ; DYNJ66410B , Lot Number:23JMB324 ; DYNJ85042, Lot Number:23JMD316 ; DYNJ68836B , Lot Number:23KLA266 ; DYNJ68836B , Lot Number:23LLB070 ; DYNJ54790F , Lot Number:23JMC258 ; DYNJ54790F , Lot Number:23LMI220 ; DYNJ61699D , Lot Number:23HBI192 ; DYNJ80232, Lot Number:24ABL377 ; DYNJ80232H , Lot Number:24ABL377 ; DYNJ42830D , Lot Number:23IBA622 ; DYNJ0200004B , Lot Number:23GMI902 ; DYNJ36100C , Lot Number:22KMH152 ; DYNJ44356I , Lot Number:22FBL880 ; DYNJ44356I , Lot Number:22HBA943 ; DYNJ44356I , Lot Number:23BBL664 ; DYNJ44356I , Lot Number:23DBC100 ; DYNJ44356I , Lot Number:23FBK101 ; DYNJ44356J , Lot Number:23HBJ438 ; DYNJ44356J , Lot Number:23JBB906 ; DYNJ44356J , Lot Number:23LBA253 ; DYNJ44356J , Lot Number:24ABQ985 ; DYNJ44356J , Lot Number:24BBC038 ; DYNJ44356J , Lot Number:24BBL641 ; DYNJ62410A , Lot Number:23GBS482 ; DYNJ62410A , Lot Number:23HBN329 ; EYT102K, Lot Number:23IDA309 ; DYNJ66003D , Lot Number:24CBE592 ; DYNJ66003C , Lot Number:21FBK406 ; DYNJ66003C , Lot Number:21CBM686 ; DYNJ66003C , Lot Number:21DBT747 ; DYNJ66003D , Lot Number:22FBC039 ; DYNJ66003D , Lot Number:23JBE968 ; DYNJ66003D , Lot Number:23JBQ750 ; DYNJ66003D , Lot Number:24ABO688 ; DYNJ66003D , Lot Number:24BBG701 ; DYNJ66004B , Lot Number:24CBE607 ; DYNJ66004A , Lot Number:20LBE742 ; DYNJ66004B , Lot Number:21FBT041 ; DYNJ66004B , Lot Number:22EBR001 ; DYNJ66004B , Lot Number:22FBS391 ; DYNJ66004B , Lot Number:24BBL458 ; DYNJ19973S , Lot Number:23JME916 ; DYNJ57258G , Lot Number:23IBB036 ; DYNJ44671C , Lot Number:21HDB625 ; DYNJ44671D , Lot Number:22ADA848 ; DYNJ44671D , Lot Number:22CDA486 ; DYNJ0559863K , Lot Number:23LMA799 ; DYNJ37682J , Lot Number:23BMD927 ; DYNJ37682K , Lot Number:23HMD991 ; DYNJ63272B , Lot Number:23IBG537 ; MNS13080 , Lot Number:21DBC890 ; DYNJ66022A , Lot Number:23HBR314 ; DYNJ66022A , Lot Number:23HBR356 ; DYNJ63638A , Lot Number:23IBE732 ; DYNJ40135C , Lot Number:23GBW482 ; DYNJ55484A , Lot Number:21KBH827 ; DYNJ55484A , Lot Number:21LBO808 ; DYNJ55484A , Lot Number:22FBJ347 ; DYNJ55484A , Lot Number:23HBM901 ; DYNJ55484A , Lot Number:23IBR023 ; DYNJ60960A , Lot Number:23BBJ999 ; DYNJ60960A , Lot Number:23CBS340 ; DYNJ60960A , Lot Number:23EBD517 ; DYNJ60960A , Lot Number:23FBI464 ; DYNJ60960A , Lot Number:23GBM127 ; DYNJ67973B , Lot Number:24ABM354 ; DYNJ83799C , Lot Number:23HBM374 ; DYNJ83799C , Lot Number:23IBE242 ; VAL013EPWEC, Lot Number:23IBD749 ; VAL038BSCLC, Lot Number:21FBI402 ; VAL038BSCLC, Lot Number:22FBI684 ; VAL038BSCLC, Lot Number:22GBI953 ; VAL038BSCLC, Lot Number:22HBE695 ; VAL038BSCLC, Lot Number:22HBH463 ; VAL038BSCLC, Lot Number:23BBP414 ; VAL038BSCLC, Lot Number:23CBA206 ; VAL038BSCLC, Lot Number:23GBN565 ; VAL038BSCLC, Lot Number:23IBQ356 ; VAL038BSCLC, Lot Number:23JBQ925 ; VAL038BSCLC, Lot Number:20IBH345 ; VAL038BSCLC, Lot Number:24ABL864 ; VAL038BSCLC, Lot Number:24BBP892 ; VAL038BSCLC, Lot Number:20LBY788 ; VAL038BSCLC, Lot Number:21CBV826 ; VAL038BSCLC, Lot Number:21FBE159 ; DYNJ56293D , Lot Number:23IDB161 ; DYNJ64051C , Lot Number:23CDC473 ; DYNJ64051C , Lot Number:23GDA852 ; CMPJ06528F , Lot Number:23KMA165 ; DYNJ38400I , Lot Number:21IMA857 ; DYNJ38400I , Lot Number:21IMH583 ; DYNJ38400I , Lot Number:22DMD856 ; DYNJ38400J , Lot Number:23KMG804 ; DYNJ60960A , Lot Number:22EMA599 ; DYNJ64284C , Lot Number:21HMA162 ; DYNJ64284C , Lot Number:23CMA702 ; DYNJ64284D , Lot Number:23FMJ780 ; DYNJ64284D , Lot Number:23KMD668 ; DYNJ64284D , Lot Number:23LMD080 ; DYNJ64284D , Lot Number:24AMG449 ; DYNJ64051D , Lot Number:23LLA491 ; VAL013EPFEG, Lot Number:23IBQ281 ; VAL013EPFEI, Lot Number:23LBH694 ; DYNJ63096A , Lot Number:23FBC693 ; DYNJ63096A , Lot Number:23GBW469 ; DYNJ0516258W , Lot Number:23KMB305 ; DYNJ0559777C , Lot Number:21HMC030 ; DYNJ0559777C , Lot Number:22CMH758 ; DYNJ0559777D , Lot Number:23EMK184 ; DYNJ0559777F , Lot Number:23GMG783 ; DYNJ0559777F , Lot Number:23JME881 ; DYNJ0559777F , Lot Number:23KMB427 ; DYNJ0559777F , Lot Number:24AMB790 ; DYNJ53097C , Lot Number:23IBE728 ; DYNJ27580F , Lot Number:22OBG462 ; DYNJ27580F , Lot Number:22FBR432 ; DYNJ27580F , Lot Number:22GBX560 ; DYNJ27580F , Lot Number:22JBE628 ; DYNJ27580F , Lot Number:22JBT093 ; DYNJ27580F , Lot Number:23BBO703 ; DYNJ27580G , Lot Number:23CBP779 ; DYNJ27580G , Lot Number:23DBS205 ; DYNJ27580G , Lot Number:23FBC630 ; DYNJ27580G , Lot Number:23GBC544 ; DYNJ27580G , Lot Number:23HBM937 ; DYNJ27580G , Lot Number:23KBQ070 ; DYNJ21627L , Lot Number:24AMF970 ; DYNJ81183A , Lot Number:22GBM466 ; DYNJ53678B , Lot Number:21LBJ879 ; DYNJ53678B , Lot Number:22FBY588 ; DYNJ53678B , Lot Number:22IBQ735 ; DYNJ53678B , Lot Number:23BBQ152 ; DYNJ53678B , Lot Number:23HBP513 ; DYNJ66038A , Lot Number:23HBT377 ; DYNJ0608576P , Lot Number:23JME914 ; DYNJ67519A , Lot Number:23HBT970 ; DYNJ67519A , Lot Number:23IBE623 ; DYNJ902885D, Lot Number:23IBD838 ; DYNJ902885D, Lot Number:23IBD935 ; DYNJ902885D, Lot Number:23IBM274 ; DYNJ54393D , Lot Number:23HBK063 ; DYNJ34288K , Lot Number:22OBB445 ; DYNJ34288K , Lot Number:22GBF074 ; DYNJ34288K , Lot Number:22HBK556 ; DYNJ34288K , Lot Number:23CBC625 ; DYNJ38109C , Lot Number:21KBT541 ; DYNJ38109C , Lot Number:21LBK050 ; DYNJ38109C , Lot Number:22IBS327 ; DYNJ38109C , Lot Number:23BBA458 ; DYNJ38109C , Lot Number:23GBF633 ; DYNJ38109C , Lot Number:23GBI442 ; DYNJ38109C , Lot Number:23IBC961 ; DYNJ67973A , Lot Number:22JBK486 ; DYNJ67973A , Lot Number:23ABM186 ; DYNJ67973A , Lot Number:23GBI608 ; DYNJ67973A , Lot Number:23IBC924 ; DYNJ55840A , Lot Number:21GMD889 ; DYNJ55840A , Lot Number:21IMC582 ; DYNJ55840A , Lot Number:22AMC428 ; DYNJ55840A , Lot Number:22CMH721 ; DYNJ55840A , Lot Number:23FMJ603 ; DYNJ55840A , Lot Number:23KMC842 ; DYNJ82201, Lot Number:23IDA107 ; DYNJ901206D, Lot Number:22HBW709 ; DYNJ901206D, Lot Number:23DBP697 ; CMPJ05506D , Lot Number:22HBH563 ; CMPJ05506D , Lot Number:23CBH848 ; DYNJ65156A , Lot Number:23KDA430 ; DYNJ47273I , Lot Number:22LMD338 ; DYNJ80556C , Lot Number:24ABL912 ; DYNJ60631B , Lot Number:23HBP777 ; DYNJ60631B , Lot Number:23HBZ160 ; DYNJVB91043A , Lot Number:23JDC086 ; DYNJVB91043A , Lot Number:23KDB345 ; CMPJ11101, Lot Number:23IDB922 ; CMPJ11101, Lot Number:23JDA637 ; DYNJ42845D , Lot Number:23IBA707 ; DYNJ44507J , Lot Number:23HBV391 ; DYNJ85449, Lot Number:24ABD671 ; DYNJ57970C , Lot Number:23CBS337 ; DYNJ57970C , Lot Number:23DBB080 ; DYNJ57970D , Lot Number:23JBT640 ; DYNJ57970D , Lot Number:23KBN028 ; DYNJ57970D , Lot Number:24ABB977 ; DYNJ57970D , Lot Number:24ABV141 ; DYNJ34159D , Lot Number:23HBV666 ; DYNJ80483, Lot Number:24ABK886 ; DYNJ66624A , Lot Number:23GMG818 ; DYNJ81824, Lot Number:23HBP061 ; DYNJ68627, Lot Number:21CBM277 ; DYNJ68627B , Lot Number:23IBC749 ; DYNJ68627B , Lot Number:23GMB879 ; DYNJ68627C , Lot Number:24AMB768 ; DYNJ43818, Lot Number:22AMD337 ; DYNJ43818, Lot Number:23BMC400 ; DYNJ43818, Lot Number:23GMI178 ; DYNJ43818, Lot Number:23JMK100 ; CMPJ21591J , Lot Number:23HBI095 ; DYNJ60021A , Lot Number:23HBT040 ; DYNJ43050N , Lot Number:23JMG418 ; CDS810023J , Lot Number:23HBV221 ; DYNJ900374N, Lot Number:22FBB171 ; DYNJ900374N, Lot Number:22FBC598 ; DYNJ900374N, Lot Number:22HBV856 ; CMPJ04908M , Lot Number:22EBF948 ; CMPJ04908M , Lot Number:22HBK544 ; DYNJ83503, Lot Number:23CMC327 ; DYNJ83503, Lot Number:23GMD280 ; DYNJ83503, Lot Number:23JMF577 ; DYNJ53853F , Lot Number:23KMB732 ; DYNJ20988G , Lot Number:21JMG468 ; DYNJ20988G , Lot Number:22AME490 ; DYNJ20988G , Lot Number:22CMA075 ; DYNJ20988G , Lot Number:22DME734 ; DYNJ20988G , Lot Number:22FMA831 ; DYNJ20988G , Lot Number:22GMC750 ; DYNJ20988G , Lot Number:22HMC747 ; DYNJ20988G , Lot Number:22IME434 ; DYNJ20988G , Lot Number:22LMD535 ; DYNJ37739L , Lot Number:23IBE128 ; DYNJ40839A , Lot Number:23KMC318 ; DYNJ40839A , Lot Number:24AMF192 ; DYNJ80114, Lot Number:21KMD689 ; DYNJ80114, Lot Number:22AMF833 ; DYNJ80114, Lot Number:22DMD369 ; DYNJ80114A , Lot Number:23BMB820 ; DYNJ80114A , Lot Number:23FMF556 ; DYNJ80114A , Lot Number:23GMD277 ; DYNJ80114A , Lot Number:23HMB645 ; DYNJ80114A , Lot Number:23HMH770 ; DYNJ80114A , Lot Number:23JMD209 ; DYNJ80114A , Lot Number:23KMB513 ; DYNJ80114A , Lot Number:23LMA560 ; VAL043VPASC, Lot Number:21LBR917 ; DYNJVB1080A, Lot Number:23IDA988 ; DYNJ55508B , Lot Number:22KBF091 ; DYNJ55508B , Lot Number:23BBI775 ; DYNJ55508B , Lot Number:23CBB250 ; DYNJ55508B , Lot Number:23EBB305 ; DYNJ55508B , Lot Number:23EBV036 ; DYNJ55508B , Lot Number:24ABW624 ; DYNJ55508B , Lot Number:24BBR693 ; DYNDA2857, Lot Number:22JME670 ; DYNDA2857A , Lot Number:23GME382 ; DYNDA2857A , Lot Number:23JMG571 ; DYNJ38495J , Lot Number:23HMG757 ; DYNJ0843936F , Lot Number:22ADB686 ; DYNJ0843936F , Lot Number:22JDA822 ; DYNJ0843936G , Lot Number:22KDB025 ; DYNJ0843936G , Lot Number:23ADA378 ; DYNJ0843936I , Lot Number:23JDA380 ; DYNJ46536F , Lot Number:21DMA451 ; DYNJ46536F , Lot Number:21FMA586 ; DYNJ46536F , Lot Number:21IMB516 ; DYNJ46536G , Lot Number:22AMD760 ; DYNJ46536G , Lot Number:22CMH930 ; DYNJ46536G , Lot Number:22DMC362 ; DYNJ46536G , Lot Number:22DMF054 ; DYNJ46536G , Lot Number:23EMJ440 ; DYNJ46536I , Lot Number:23GMI800 ; DYNJ82153, Lot Number:22ILA613 ; DYNJ82153, Lot Number:23DLA980 ; DYNJ82153, Lot Number:24BLB069
Worldwide