FDA Enforcement Class II Terminated

The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

Recall: Z-2763-2019 · Reported October 9, 2019

Enforcement

Recall Number
Z-2763-2019
Event ID
83163
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 9, 2019
Initiation Date
September 18, 2018
Classification Date
September 30, 2019
Termination Date
February 21, 2024
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

Reason

Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).

Code Info

Lot Number 831755.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.

Quantity

Access Cortisol Reagent Assay