FDA Enforcement Class II Terminated

Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluorescence

Recall: Z-2760-2017 · Reported August 9, 2017

Enforcement

Recall Number
Z-2760-2017
Event ID
77759
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Olympus Scientific Solutions Americas
Voluntary / Mandated
FDA Mandated
Initial Notification
Letter
Report Date
August 9, 2017
Initiation Date
July 11, 2017
Classification Date
August 2, 2017
Termination Date
May 3, 2023
Address
48 Woerd Ave, N/A, Waltham, MA, 02453-3824, United States

Description

Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluorescence

Reason

Instruments may, under certain specific circumstances listed below, fail to stop when a test is initiated and no sample is present (proximity shut down fails to end test).

Code Info

Models : VCR and VMR.

Distribution

Worldwide Distribution - USA (nationwide) Distribution

Quantity