FDA Enforcement Class II Terminated

Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided for with a vinyl female luer connector bonded proximally and a vinyl cuff bonded distally. The cuff is pre-molded to its inflated configuration and is purposely positioned directly over a hole generated in the extruded tube for purposes of receiving the infused cardioplegia solution. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.

Recall: Z-2757-2016 · Reported September 14, 2016

Enforcement

Recall Number
Z-2757-2016
Event ID
74934
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Sorin Group USA, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 14, 2016
Initiation Date
August 12, 2016
Classification Date
September 8, 2016
Termination Date
March 27, 2017
Address
14401 W 65th Way, Arvada, CO, 80004-3503, United States

Description

Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided for with a vinyl female luer connector bonded proximally and a vinyl cuff bonded distally. The cuff is pre-molded to its inflated configuration and is purposely positioned directly over a hole generated in the extruded tube for purposes of receiving the infused cardioplegia solution. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.

Reason

The tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.

Code Info

Lot Numbers: 1425900137, 1433900079, 1433900080, 1435300088.

Distribution

Worldwide Distribution - US including IL, MO, TX. Internationally to Canada and Italy.

Quantity

1205 units worldwide