FDA Enforcement
Class II
Terminated
BD 3ml Syringe Luer-Lok" Tip Catalog 309657 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
Recall: Z-2751-2017
·
Reported July 26, 2017
Enforcement
- Recall Number
- Z-2751-2017
- Event ID
- 77513
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Becton Dickinson & Company
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 26, 2017
- Initiation Date
- September 16, 2016
- Classification Date
- July 19, 2017
- Termination Date
- March 20, 2018
- Address
- 1 Becton Dr, N/A, Franklin Lakes, NJ, 07417-1815, United States
Description
BD 3ml Syringe Luer-Lok" Tip Catalog 309657 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
Reason
BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required of 10-6.
Code Info
6061746 2021-02 6082982 2021-02 6089821 2021-03 6117858 2021-04 6117863 2021-04 6117889 2021-04 6117891 2021-04 6146962 2021-04 6173697 2021-05 6173767 2021-06 6173779 2021-06 6208614 2021-07
Distribution
Nationwide Distribution
Quantity
648,000 units