FDA Enforcement
Class II
Terminated
DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US
Recall: Z-2747-2020
·
Reported August 12, 2020
Enforcement
- Recall Number
- Z-2747-2020
- Event ID
- 85980
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 12, 2020
- Initiation Date
- June 30, 2020
- Classification Date
- July 31, 2020
- Termination Date
- November 17, 2020
- Address
- 500 Gbc Dr, Newark, DE, 19702-2466, United States
Description
DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US
Reason
If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.
Code Info
DCA Vantage Handheld Barcode Scanner #11306432 Model: DS4308 - HC0015BZZWW
Distribution
Nationwide
Quantity
1800 units