Remisol Advance running Software Version 1.12 The Normand Info Remisol Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS &). It can also be connected to a host system.

Recall: Z-2743-2020 · Reported August 12, 2020

Enforcement

Recall Number
Z-2743-2020
Event ID
85693
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Normand-Info S.A.S.U.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 12, 2020
Initiation Date
April 30, 2020
Classification Date
July 31, 2020
Termination Date
February 10, 2022
Address
7 rue Frederic Degeorge, Arras CEDEX, N/A, France

Description

Remisol Advance running Software Version 1.12 The Normand Info Remisol Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS &). It can also be connected to a host system.

Reason

Results from repeated run for WBC (White Bloodcell Count), UWBC (Urine Whit Bloodcell Count) and PLT (Platelets) are deleted in RADV. Use may lead to erroneous results.

Code Info

All serial numbers

Distribution

Worldwide distribution: US (nationwide) including states of: CA, GA, IL, IN, KS, MI, MO, NY, NC, OH, PA, SD, TN, VA, and WV; and countries of: Bahrain, Belgium, Canada, China, Czech Republic, France, Germany, Hong Kong, Israel, Malaysia, Netherlands, Switzerland, Taiwan,Turkey, and United Kingdom.

Quantity

663