FDA Enforcement
Class II
Ongoing
Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease Model Number: 00-1300-6053
Recall: Z-2742-2020
·
Reported August 5, 2020
Enforcement
- Recall Number
- Z-2742-2020
- Event ID
- 86027
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- OrthoPediatrics Corp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 5, 2020
- Initiation Date
- July 6, 2020
- Classification Date
- July 30, 2020
- Address
- 2850 Frontier Dr, N/A, Warsaw, IN, 46582-7001, United States
Description
Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease Model Number: 00-1300-6053
Reason
Product may have incorrect laser etched alignment lines, the non-conforming laser etch may have compromised strength, lead to unintended fracture of the device
Code Info
Lot Number: 108332-A UDI: (01)0084113212400(10)108332A
Distribution
US Nationwide distribution including in the states of Al, CA, FL, GA, KY, MI, MO, NC, NY, OH, TX.
Quantity
73 units