Terminated
2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plating System for trauma and reconstructive procedures of small bones.
Recall: Z-2738-2015
·
Reported September 16, 2015
Enforcement
- Recall Number
- Z-2738-2015
- Event ID
- 71936
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Trilliant Surgical Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 16, 2015
- Initiation Date
- July 31, 2015
- Classification Date
- September 4, 2015
- Termination Date
- January 26, 2017
- Address
- 6721 Portwest Dr Ste 160, N/A, Houston, TX, 77024-8019, United States
Description
2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plating System for trauma and reconstructive procedures of small bones.
Reason
Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.
Code Info
TSL002605
Distribution
NY, OH, CA, SC, NJ, AZ, VA, IN, FL, MN, OK, AR, MO, WI, GA, SD, IA, IL, TX
Quantity
108 units