FDA Enforcement Class II Terminated

Waterpik Sonic-Fusion Professional, Model: SF-02 Product Usage: The device is composed of a unit base and an oral irrigator/ battery operated brush power handle. The device provides a pulsating stream of potable water and/or powered toothbrush action to remove plaque and food particles from between the teeth and below the gum line. The power handle can also function as a rechargeable battery powered mechanical toothbrush.

Recall: Z-2734-2018 · Reported August 15, 2018

Enforcement

Recall Number
Z-2734-2018
Event ID
80474
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Water Pik, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 15, 2018
Initiation Date
July 2, 2018
Classification Date
August 8, 2018
Termination Date
May 24, 2022
Address
1730 East Prospect Road, N/A, Fort Collins, CO, 80553-0001, United States

Description

Waterpik Sonic-Fusion Professional, Model: SF-02 Product Usage: The device is composed of a unit base and an oral irrigator/ battery operated brush power handle. The device provides a pulsating stream of potable water and/or powered toothbrush action to remove plaque and food particles from between the teeth and below the gum line. The power handle can also function as a rechargeable battery powered mechanical toothbrush.

Reason

Charging base may overheat with localized melting and sparking, possibly causing fire, shock or burns.

Code Info

Serial #: SF02 17 06 01 through SF02 18 06 28

Distribution

U.S. Nationwide Distribution in the states of TX, KY, WA, NY, VA, CA, NC, IA, MA, OR, OH, GA, NV, CT, IL, SC, PA, MD, IN, WV, ND, MS, CO, AZ, WI, KS, FL, HI, OK, VT, MO, ID, RI, MN, MI, AR, LA, TN, DC, NJ, NH, NE, AL, UT, ME, NM, AK, DE, SD, MT, WY and country of Canada

Quantity

3797