FDA Enforcement Class II Terminated

Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.

Recall: Z-2733-2015 · Reported September 9, 2015

Enforcement

Recall Number
Z-2733-2015
Event ID
71878
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 9, 2015
Initiation Date
July 28, 2015
Classification Date
September 3, 2015
Termination Date
April 13, 2016
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.

Reason

Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL.

Code Info

Catalogue number: 33860 Lot numbers: 431990 434727 430969 522896 523251 Expiration date: 10/31/16 12/31/16 08/31/16 04/30/17 05/31/17

Distribution

Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Algeria, Australia, Austria, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Republic of, Lebanon, Lithuania, Mexico, Morocco, New Zealand, Oman, Philippines, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam and Yemen.

Quantity

3,035 (U.S.) and 7,123 (International)